Hand-Drilled External Fixator Pins Have Superior Biomechanical Strength Compared to Power-Drilling
Military Medicine · 21.07.2026
DOI: 10.1093/milmed/usag330 ↗Kurz gefasst
Fachartikel in der Zeitschrift Military Medicine, online erschienen am 21.07.2026. Verfasst von Mason H Adams und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Brooke Army Medical Center, San Antonio, TX 78234, United States. Die Arbeit nennt 10 Literaturangaben.
Abstract (Originalsprache)
Objectives
In austere environments, surgeons commonly use a manual brace instead of a power drill to insert external fixation pins. Yet the biomechanical performance of manually inserted pins is uncertain. This study aimed to compare the pull-out strength of external fixator pins inserted with a manual brace with that of pins inserted with a power drill. We hypothesized that manual insertion would result in reduced pull-out strength compared to powered methods. Materials and Methods Self-drilling, self-tapping external fixator pins were inserted into 90 porcine femoral shaft segments using 3 techniques: hand drilled, power drilled, and pre-drilled/power-inserted. Each of the 3 techniques was performed by 3 surgeons, with a total of 30 specimens per technique and equal distribution between diaphyseal and metaphyseal bone. Pull-out strength to failure was measured using a materials testing machine. A generalized linear mixed effects model (GLMM) (least squares means [LMS] ± standard error) and post hoc pairwise comparisons were utilized for statistical analysis.
Results
A GLMM identified drilling technique (P = .02) and diaphyseal pin location (P = .03) as significant predictors of pin pull-out strength, while the performing surgeon had no significant effect. Contrary to the hypothesis, post hoc pairwise analysis demonstrated hand drilled pins (LMS: 1,542.6 N ± 74.7) had significantly greater pull-out strength compared to both power drilled (LMS: 1,344.2 N ± 74.1) and pre-drilled/power-inserted pins (LMS: 1,248.5 N ± 89.8) (P = .05 and .006, respectively). No significant difference was observed between the power drilled and pre-drilled power-inserted techniques.
Conclusions
The pull-out strength of pins inserted with a hand brace was superior to that of pins inserted with power tools. This finding challenges common assumptions about the mechanical inferiority of manual drilling. These findings support the continued use and funding of hand drilled external fixator training programs within military and humanitarian aid settings. Finally, pre-drilling does not appear to confer any mechanical advantage over direct, powered pin insertion.
Autor:innen und Institutionen
- Department of Orthopedic Surgery, Brooke Army Medical Center, San Antonio, TX 78234, United States.
- United States Army Institute of Surgical Research, Fort Sam Houston, San Antonio, TX 78234, United States.
- 59MDW/CC, 1100 Willford Hall Loop, Lackland AFB, TX 78236, United States.
- Uniformed Services University of the Health Sciences, Bethesda, MD 20814, United States.
ORCID-Profile: Simon Milne
Zitieren
APA
Adams, M. H., Carmack, D. B., Bowers, Z. D., Milne, S., Bates, T., Moritz, R., Mason, G. B., Gaeth, C., & Cognetti, D. J. (2026). Hand-Drilled External Fixator Pins Have Superior Biomechanical Strength Compared to Power-Drilling. Military Medicine. https://doi.org/10.1093/milmed/usag330
Vancouver
Adams MH, Carmack DB, Bowers ZD, Milne S, Bates T, Moritz R, et al. Hand-Drilled External Fixator Pins Have Superior Biomechanical Strength Compared to Power-Drilling. Military Medicine. 2026. doi:10.1093/milmed/usag330
BibTeX
@article{Adams2026,
author = {Adams, Mason H and Carmack, David B and Bowers, Zachary D and Milne, Simon and Bates, Taylor and Moritz, Robert and Mason, Gabriel B and Gaeth, Catharina and Cognetti, Daniel J},
title = {Hand-Drilled External Fixator Pins Have Superior Biomechanical Strength Compared to Power-Drilling},
journal = {Military Medicine},
year = {2026},
doi = {10.1093/milmed/usag330}
}
EUVIMED