Interference of hemoglobin variants with HbA1c measurements by six commonly used HbA1c methods

Mingyang Li, Song Ge, Xin Shu, Xiongjun Wu, Haiyan Liu, Anping Xu iD, Ling Ji

Laboratory Medicine · 04.11.2024 · 55 (6) · 708-712

DOI: 10.1093/labmed/lmae034 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laboratory Medicine, online erschienen am 27.05.2024. Verfasst von Mingyang Li und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, China. Hauptthemen laut MEDLINE: Glycated Hemoglobin. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 18 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Glycated hemoglobin, or hemoglobin A1c (HbA1c), serves as a crucial marker for diagnosing diabetes and monitoring its progression. We aimed to assess the interference posed by common Hb variants on popular HbA1c measurement systems.

Methods

A total of 63 variant and nonvariant samples with target values assigned by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)-calibrated methods were included. We assessed 6 methods for measuring HbA1c in the presence of HbS, HbC, HbD, HbE, and fetal hemoglobin (HbF): 2 cation-exchange high-performance liquid chromatography (HPLC) methods (Bio-Rad D-100 and HLC-723 G8), a capillary electrophoresis (CE) method (Sebia Capillarys 3 TERA), an immunoassay (Roche c501), an enzyme assay system (Mindray BS-600M), and a boronate affinity method (Primus Premier Hb9210).

Results

The HbA1c results for nonvariant samples from the 6 methods were in good agreement with the IFCC-calibrated method results. The Bio-Rad D-100, Capillarys 3, Mindray BS-600M, Premier Hb9210, and Roche c501 showed no interference from HbS, HbC, HbD, and HbE. Clinically significant interference was observed for the HLC-723 G8 standard mode. Elevated HbF levels caused significant negative biases for all 6 methods, which increased with increasing HbF concentration.

Conclusion

Elevated levels of HbF can severely affect HbA1c measurements by borate affinity, immunoassays, and enzyme assays.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Glycated HemoglobinHumansChromatography, High Pressure LiquidElectrophoresis, CapillaryHemoglobins, AbnormalImmunoassayBlood Chemical Analysis

Stichwörter der Autor:innen

HPLCboronate affinityelectrophoresisenzymefetal hemoglobinhemoglobin Alchemoglobin variantimmunoassay

Substanzen

Glycated HemoglobinHemoglobins, Abnormalhemoglobin A1c protein, human

Autor:innen und Institutionen

  • Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, China.
  • Department of Laboratory Medicine, Wuhan Asia General Hospital Affiliated to Wuhan University of Science and Technology, Wuhan, China.
  • Department of Laboratory Medicine, Shenzhen Integrated Traditional Chinese and Western Medicine Hospital, Shenzhen, China.

ORCID-Profile: Anping Xu

Zitieren

APA

Li, M., Ge, S., Shu, X., Wu, X., Liu, H., Xu, A., & Ji, L. (2024). Interference of hemoglobin variants with HbA1c measurements by six commonly used HbA1c methods. Laboratory Medicine, 55(6), 708-712. https://doi.org/10.1093/labmed/lmae034

Vancouver

Li M, Ge S, Shu X, Wu X, Liu H, Xu A, et al. Interference of hemoglobin variants with HbA1c measurements by six commonly used HbA1c methods. Laboratory Medicine. 2024;55(6):708-712. doi:10.1093/labmed/lmae034

BibTeX
@article{Li2024,
  author = {Li, Mingyang and Ge, Song and Shu, Xin and Wu, Xiongjun and Liu, Haiyan and Xu, Anping and Ji, Ling},
  title = {Interference of hemoglobin variants with HbA1c measurements by six commonly used HbA1c methods},
  journal = {Laboratory Medicine},
  year = {2024},
  volume = {55},
  number = {6},
  pages = {708--712},
  doi = {10.1093/labmed/lmae034}
}
Kostenlos starten →