Effect of tumor size on motion tracking and dosimetric accuracy using Radixact Synchrony

Ryoto Kaido iD, Akihiro Takemura iD, Kakeru Sato, Naoki Kinoshita iD, Kouhei Sasamoto, Tomohiro Kotsuji, Hiroki Shioura, Toshiki Tateishi iD, Tetsuya Tsujikawa iD

Journal of Radiation Research · 2026

DOI: 10.1093/jrr/rrag002 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Radiation Research, 2026. Verfasst von Ryoto Kaido und 8 weiteren Autor:innen. Beteiligte Einrichtung: Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa Prefecture 920-0942, Jap…. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Tumor Burden, Radiometry. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aimed to evaluate how planning target volume (PTV) size influences treatment-planning parameters and their relationship to tracking and irradiation accuracy. A high-resolution two-dimensional diode array mounted on a motion phantom was irradiated under four respiratory waveforms using Fiducial Tracking with Respiratory Correlation mode. Spherical PTVs with diameters of 1–4 cm were defined. Treatment-planning parameters, including beam-on time, couch speed, gantry period, kilovoltage imaging interval and modulation factor, were analyzed to assess their relationship with tracking and irradiation accuracy. The root-mean-square error (${\delta}_{\mathrm{RMS}}$) between tracked and reference target positions, γ pass rate and potential difference as the control parameter of the correlation model were compared across PTV sizes. Correlation and partial correlation analyses were performed to evaluate the relationships between the ${\delta}_{\text{RMS}}$ and treatment-planning parameters. The ${\delta}_{\text{RMS}}$ was influenced by the inherent interdependence of beam-on time, gantry period and the kV imaging interval. The kV imaging interval for the PTV of φ1 cm was 4.2 s, longer than those for the other PTV sizes, which made the potential difference unstable during the initial tracking period and kept γ pass rates below 92% across all respiratory waveforms. For the PTVs of φ2–3 cm, potential difference stabilized within ~20 s after beam-on and was associated with an improvement in the γ pass rate. In contrast, for the PTV of φ4 cm, degradation of tracking accuracy was reflected in a reduction in the γ pass rate. Optimizing treatment-planning parameters for each PTV size was essential for precise motion-synchronized irradiation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedTumor BurdenRadiometryHumansPhantoms, ImagingReproducibility of ResultsMovementRadiotherapy DosageMotionSensitivity and Specificity

Stichwörter der Autor:innen

Radixact SynchronySBRTrespiratory trackingtomotherapy

Autor:innen und Institutionen

  • Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa Prefecture 920-0942, Japan.
  • Radiological Center, University of Fukui Hospital, 23-3 Matsuoka Shimoaizuki, Eiheiji-cho, Yoshida-gun, Fukui Prefecture 910-1193, Japan.
  • Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa Prefecture 920-0942, Japan.
  • Department of Radiological Sciences, Faculty of Medical Science Technology, Morinomiya University of Medical Sciences, 1-26-16 Nankokita, Suminoe-ku, Osaka Prefecture 559-8611, Japan.
  • Department of Radiology, Faculty of Medical Sciences, University of Fukui, 23-3 Matsuoka Shimoaizuki, Eiheiji-cho, Yoshida-gun, Fukui Prefecture 910-1193, Japan.

ORCID-Profile: Ryoto Kaido, Akihiro Takemura, Naoki Kinoshita, Toshiki Tateishi, Tetsuya Tsujikawa

Zitieren

APA

Kaido, R., Takemura, A., Sato, K., Kinoshita, N., Sasamoto, K., Kotsuji, T., Shioura, H., Tateishi, T., & Tsujikawa, T. (2026). Effect of tumor size on motion tracking and dosimetric accuracy using Radixact Synchrony. Journal of Radiation Research. https://doi.org/10.1093/jrr/rrag002

Vancouver

Kaido R, Takemura A, Sato K, Kinoshita N, Sasamoto K, Kotsuji T, et al. Effect of tumor size on motion tracking and dosimetric accuracy using Radixact Synchrony. Journal of Radiation Research. 2026. doi:10.1093/jrr/rrag002

BibTeX
@article{Kaido2026,
  author = {Kaido, Ryoto and Takemura, Akihiro and Sato, Kakeru and Kinoshita, Naoki and Sasamoto, Kouhei and Kotsuji, Tomohiro and Shioura, Hiroki and Tateishi, Toshiki and Tsujikawa, Tetsuya},
  title = {Effect of tumor size on motion tracking and dosimetric accuracy using Radixact Synchrony},
  journal = {Journal of Radiation Research},
  year = {2026},
  doi = {10.1093/jrr/rrag002}
}
Kostenlos starten →