Effect of neutron beam properties on dose distributions in a water phantom for boron neutron capture therapy

Akihisa Ishikawa, Hiroki Tanaka, Satoshi Nakamura, Hiroaki Kumada, Yoshinori Sakurai, Kenichi Watanabe, Sachiko Yoshihashi, Yuki Tanagami, Akira Uritani, Yoshiaki Kiyanagi

Journal of Radiation Research · 03.12.2024 · 65 (6) · 765-775

DOI: 10.1093/jrr/rrae076 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Radiation Research, online erschienen am 04.10.2024. Verfasst von Akihisa Ishikawa und 9 weiteren Autor:innen. Beteiligte Einrichtung: Research Group for Nuclear Sensing, Nuclear Science and Engineering Center, Japan Atomic Energy Agency, 2-4 Shirakata, Tokai-mura, Ibaraki 319-1195, …. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy, Phantoms, Imaging, Neutrons, Water. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

From the viewpoints of the advantage depths (ADs), peak tumor dose and skin dose, we evaluated the effect on the dose distribution of neutron beam properties, namely the ratio between thermal and epithermal neutron fluxes (thermal/epithermal ratio), fast neutron component and γ-ray component. Several parameter surveys were conducted with respect to the beam properties of neutron sources for boron neutron capture therapy assuming boronophenylalanine as the boron agent using our dose calculation tool, called SiDE. The ADs decreased by 3% at a thermal/epithermal ratio of 20–30% compared with the current recommendation of 5%. The skin dose increased with the increasing thermal/epithermal ratio, reaching a restricted value of 14 Gyeq at a thermal/epithermal ratio of 48%. The fast neutron component was modified using two different models, namely the ‘linear model’, in which the fast neutron intensity decreases log-linearly with the increasing neutron energy, and the ‘moderator thickness (MT) model’, in which the fast neutron component is varied by adjusting the MT in a virtual beam shaping assembly. Although a higher fast neutron component indicated a higher skin dose, the increment was <10% at a fast neutron component of <1 × 10−12 Gy cm2 for both models. Furthermore, in the MT model, the epithermal neutron intensity at a fast neutron component of 6.8 × 10−13 Gy cm2 was 41% higher compared with that of 2 × 10−13 Gy cm2. The γ-ray component also caused no significant disadvantages up to several times larger compared with the current recommendation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Boron Neutron Capture TherapyPhantoms, ImagingNeutronsWaterRadiotherapy DosageHumansDose-Response Relationship, RadiationSkin

Stichwörter der Autor:innen

beam propertiesboron neutron capture therapyfast neutron componentskin dosethermal/epithermal ratioγ-ray component

Substanzen

Water

Autor:innen und Institutionen

  • Research Group for Nuclear Sensing, Nuclear Science and Engineering Center, Japan Atomic Energy Agency, 2-4 Shirakata, Tokai-mura, Ibaraki 319-1195, Japan.
  • Department of Applied Energy, Graduate School of Engineering, Nagoya University, Furo-cho, Chikusa-ku, Nagoya, Aichi 464-8603, Japan.
  • Integrated Radiation and Nuclear Science, Kyoto University, 2, Asashiro-Nishi, Kumatori-cho, Sennan-gun, Osaka 590-0494, Japan.
  • Division of Radiation Safety and Quality Assurance, National Cancer Center Hospital, 5-1-1, Tsukiji, Chuo-ku, Tokyo 104-0045, Japan.
  • Proton Medical Research Center, Institute of Medicine, University of Tsukuba, 1-1-1, Tennodai, Tsukuba, Ibaraki 305-8575, Japan.
  • Department of Applied Quantum Physics and Nuclear Engineering, Kyushu University, 744, Motooka, Nishi-ku, Fukuoka 819-0395, Japan.

Zitieren

APA

Ishikawa, A., Tanaka, H., Nakamura, S., Kumada, H., Sakurai, Y., Watanabe, K., Yoshihashi, S., Tanagami, Y., Uritani, A., & Kiyanagi, Y. (2024). Effect of neutron beam properties on dose distributions in a water phantom for boron neutron capture therapy. Journal of Radiation Research, 65(6), 765-775. https://doi.org/10.1093/jrr/rrae076

Vancouver

Ishikawa A, Tanaka H, Nakamura S, Kumada H, Sakurai Y, Watanabe K, et al. Effect of neutron beam properties on dose distributions in a water phantom for boron neutron capture therapy. Journal of Radiation Research. 2024;65(6):765-775. doi:10.1093/jrr/rrae076

BibTeX
@article{Ishikawa2024,
  author = {Ishikawa, Akihisa and Tanaka, Hiroki and Nakamura, Satoshi and Kumada, Hiroaki and Sakurai, Yoshinori and Watanabe, Kenichi and Yoshihashi, Sachiko and Tanagami, Yuki and Uritani, Akira and Kiyanagi, Yoshiaki},
  title = {Effect of neutron beam properties on dose distributions in a water phantom for boron neutron capture therapy},
  journal = {Journal of Radiation Research},
  year = {2024},
  volume = {65},
  number = {6},
  pages = {765--775},
  doi = {10.1093/jrr/rrae076}
}
Kostenlos starten →