EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Cardiac Troponin T Stability Study for Regulatory Approval of the 6th Generation High-Sensitivity Assay

Nina M Diederiks iD, Jorik H Amesz iD, Fred P H T M Romijn, Enrico de Koning, Mark J Boogers, Christa M Cobbaert

The Journal of Applied Laboratory Medicine · 26.09.2026

DOI: 10.1093/jalm/jfag148 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 26.09.2026. Verfasst von Nina M Diederiks und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Roche Diagnostics.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

High-sensitivity cardiac troponin T (hs-cTnT) assays must meet strict criteria prior to regulatory approval, as the test is central to myocardial infarction diagnosis. Evaluating analyte stability is essential for ensuring accurate results in stored samples. Here, cTnT stability using the novel sixth-generation assay was evaluated. We also discuss regulatory complexity.

Methods

From patients with acute coronary syndrome, serum and lithium heparin plasma (Li-Hep) were collected, pooled per matrix, and aliquoted. One aliquot was immediately analyzed (fresh) while the remainder was stored at ≤−60°C. Selected aliquots were thawed and measured (fresh-frozen) before transfer to 15–25°C for 3 and 5 h, or 2–8°C for 11 and 24 h. Long-term aliquots were transferred frozen to −15 to −25°C for ≥13 months or ≤−60 for 9 and ≥13 months. Measurements were performed in quadruplicate using the Elecsys® hs-cTnT Gen 6 assay on a Cobas® Pure analyzer (Roche Diagnostics). Recovery was calculated relative to fresh measurements. Maximum permissible differences (MPD) were ±1.2 ng/L, ±20%, and ±15% for cTnT concentrations <6 ng/L, ≥6–≤24 ng/L, and >24 ng/L, respectively, allowing one exceedance per condition.

Results

In total, >2000 measurements were performed in 18 distinct samples per matrix. Median cTnT concentrations were 49.3 ng/L (8.4–782) in Li-Hep and 49.8 ng/L (8.5–822) in serum. The observed deviation in recovery was below the MPD for all storage conditions and acceptance criteria were met.

Conclusions

Stability of cTnT using the sixth-generation hs-cTnT test was demonstrated across different storage conditions for serum and Li-Hep.

Autor:innen und Institutionen

  • Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands.
  • Department of Cardiology, Leiden University Medical Center, Leiden, the Netherlands.

ORCID-Profile: Nina M Diederiks, Jorik H Amesz

Förderung

  • Roche Diagnostics

Zitieren

APA

Diederiks, N. M., Amesz, J. H., Romijn, F. P. H. T. M., de Koning, E., Boogers, M. J., & Cobbaert, C. M. (2026). Cardiac Troponin T Stability Study for Regulatory Approval of the 6th Generation High-Sensitivity Assay. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfag148

Vancouver

Diederiks NM, Amesz JH, Romijn FPHTM, de Koning E, Boogers MJ, Cobbaert CM. Cardiac Troponin T Stability Study for Regulatory Approval of the 6th Generation High-Sensitivity Assay. The Journal of Applied Laboratory Medicine. 2026. doi:10.1093/jalm/jfag148

BibTeX
@article{Diederiks2026,
  author = {Diederiks, Nina M and Amesz, Jorik H and Romijn, Fred P H T M and de Koning, Enrico and Boogers, Mark J and Cobbaert, Christa M},
  title = {Cardiac Troponin T Stability Study for Regulatory Approval of the 6th Generation High-Sensitivity Assay},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2026},
  doi = {10.1093/jalm/jfag148}
}
Kostenlos starten →