EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Ten-Year Reintervention-Free Survival After Surgical Aortic Valve Replacement for Different Prosthetic Aortic Valve Models: Comparison Using Real-World Data From the Netherlands Heart Registration

Hector W L de Beaufort iD, Maaike M Roefs iD, Edgar J Daeter, Wilson W L Li, Wim Jan P van Boven, Niels P van der Kaaij iD, Dennis van Veghel, Cardiothoracic Surgery Registration Committee of the Netherlands Heart Registration

Interdisciplinary CardioVascular and Thoracic Surgery · 03.11.2026 · 41 (11)

DOI: 10.1093/icvts/ivag240 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Interdisciplinary CardioVascular and Thoracic Surgery, online erschienen am 31.08.2026. Verfasst von Hector W L de Beaufort und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiothoracic Surgery, St. Antonius Hospital, 3435 CM Nieuwegein, The Netherlands. Hauptthemen laut MEDLINE: Heart Valve Prosthesis Implantation, Aortic Valve, Heart Valve Prosthesis, Reoperation. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To compare long-term outcomes across different models of mechanical and biological aortic valve prostheses after surgical aortic valve replacement (SAVR) in the Netherlands.

Methods

Data from the Netherlands Heart Registration (NHR) on isolated, primary SAVR procedures performed from 2007 to 2018 in adult patients were used, excluding active endocarditis. Ten-year survival and freedom from valve-related reintervention were estimated for models that were implanted >200 times, using Cox proportional hazards models for survival and Fine and Gray’s models with death as the competing event for freedom of reintervention.

Results

For 11 139 patients treated with a biological prosthetic aortic valve, 10-year survival was highest for the Perimount Magna Ease (72.6%) and Epic (70.0%), and significantly lower for the Perimount Magna (65.1%), Trifecta (63.1%), Perimount (61.4%), Freedom Solo (51.5%), and Mitroflow (50.7%). Ten-year freedom from reintervention was highest for the Perimount Magna (99.3%), and significantly lower for the Perimount Magna Ease (98.2%), Perimount (98.2%), Epic (97.3%), Trifecta (96.8%), Freedom Solo (93.2%), and Mitroflow (90.5%). For 2123 patients treated with a mechanical prosthetic aortic valve, 10-year survival was highest for the Regent (91.2%) and not significantly different for the Open Pivot (89.2%), Slimline (87.5%), HP (87.3%), and significantly lower for the Carbomedics Reduced (87.0%). Ten-year freedom from reintervention was highest for the Carbomedics Reduced (97.1%), not significantly different for the Open Pivot (96.1%), Slimline (96.5%), Regent (95.9%), and significantly lower for the HP (94.9%).

Conclusions

Differences in 10-year outcomes after SAVR were observed for the different prosthetic aortic valve models that were used in the NHR, especially for biological prosthetic aortic valve models, with survival varying from 50.7% to 72.6% and freedom from reintervention from 90.5% to 99.3%.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Valve Prosthesis ImplantationAortic ValveHeart Valve ProsthesisReoperationBioprosthesisAortic Valve StenosisPostoperative ComplicationsHumansNetherlandsRegistriesFemaleAgedMaleProsthesis DesignTime FactorsMiddle AgedRisk Factors

Stichwörter der Autor:innen

aortic valve replacementlong-term outcomesregistry

Autor:innen und Institutionen

  • Department of Cardiothoracic Surgery, St. Antonius Hospital, 3435 CM Nieuwegein, The Netherlands.
  • Netherlands Heart Registration, 3511 EP Utrecht, The Netherlands.
  • Department of Cardiothoracic Surgery, Radboud University Medical Center, 6525 GA Nijmegen, The Netherlands.
  • Department of Cardiothoracic Surgery, Amsterdam University Medical Center,1105 AZ Amsterdam, The Netherlands.
  • Department of Cardiothoracic Surgery, Erasmus Medical Center, 3000 CA Rotterdam, The Netherlands.

ORCID-Profile: Hector W L de Beaufort, Maaike M Roefs, Niels P van der Kaaij

Zitieren

APA

de Beaufort, H. W. L., Roefs, M. M., Daeter, E. J., Li, W. W. L., van Boven, W. J. P., van der Kaaij, N. P., van Veghel, D., & Cardiothoracic Surgery Registration Committee of the Netherlands Heart Registration (2026). Ten-Year Reintervention-Free Survival After Surgical Aortic Valve Replacement for Different Prosthetic Aortic Valve Models: Comparison Using Real-World Data From the Netherlands Heart Registration. Interdisciplinary CardioVascular and Thoracic Surgery, 41(11). https://doi.org/10.1093/icvts/ivag240

Vancouver

de Beaufort HWL, Roefs MM, Daeter EJ, Li WWL, van Boven WJP, van der Kaaij NP, et al. Ten-Year Reintervention-Free Survival After Surgical Aortic Valve Replacement for Different Prosthetic Aortic Valve Models: Comparison Using Real-World Data From the Netherlands Heart Registration. Interdisciplinary CardioVascular and Thoracic Surgery. 2026;41(11). doi:10.1093/icvts/ivag240

BibTeX
@article{deBeaufort2026,
  author = {de Beaufort, Hector W L and Roefs, Maaike M and Daeter, Edgar J and Li, Wilson W L and van Boven, Wim Jan P and van der Kaaij, Niels P and van Veghel, Dennis and Cardiothoracic Surgery Registration Committee of the Netherlands Heart Registration, Cardiothoracic Surgery Registration Committee of the Netherlands Heart Registration},
  title = {Ten-Year Reintervention-Free Survival After Surgical Aortic Valve Replacement for Different Prosthetic Aortic Valve Models: Comparison Using Real-World Data From the Netherlands Heart Registration},
  journal = {Interdisciplinary CardioVascular and Thoracic Surgery},
  year = {2026},
  volume = {41},
  number = {11},
  doi = {10.1093/icvts/ivag240}
}
Kostenlos starten →