EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Initial Post-Commercialization Experience Using a Thoracic Branch Endoprosthesis: Broad Application to Real-World Patients

Xiaoying Lou iD, Patrick R Vargo, Francis Caputo, Sean Lyden, Benjamin Kramer, Marijan Koprivanac, Vidyasagar Kalahasti, Milind Desai iD, Eric E Roselli iD, Aorta Center Investigators

European Journal of Cardio-Thoracic Surgery · 2026

DOI: 10.1093/ejcts/ezaf452 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Journal of Cardio-Thoracic Surgery, 2026. Verfasst von Xiaoying Lou und 9 weiteren Autor:innen. Beteiligte Einrichtung: Aorta Center, Heart, Vascular, and Thoracic Institute, Cleveland Clinic, Cleveland, OH 44195, United States. Hauptthemen laut MEDLINE: Endovascular Procedures, Blood Vessel Prosthesis, Aortic Aneurysm, Thoracic, Blood Vessel Prosthesis Implantation. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: High-Risk Cardiovascular Research Philanthropic Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The first thoracic branched endoprosthesis (TBE) device for arch and proximal descending aortic thoracic endovascular aortic repair (TEVAR) became commercially available in the United States in 2022 (WL Gore, Flagstaff, AZ, United States). The pivotal approval study was limited to zone 2 deployment in cases of aortic aneurysm, but its use in aortic dissections and deployment in zones 0 and 1 remain investigational. We describe early results of Gore TBE device deployment in real-world patients and in a variety of clinical settings including off-label indications.

Methods

Patient characteristics, procedural details, and postoperative outcomes were assessed in this retrospective single-centre cohort study. Twenty-one patients (38.2%) were not open surgical candidates due to comorbidities. Follow-up evaluations included computed tomography angiography imaging of the device and its branches in addition to out Patient evaluation at discharge, 3 and 12 months, and then annually thereafter. The rate of endoleaks and re-interventions were assessed over the course of follow-up.

Results

Between September 2022 and September 2024, 55 consecutive patients underwent aortic repair using the Gore TBE endoprosthesis at our institution. Ten (18.2%) cases were performed in an urgent/emergency setting. Indication for TBE repair was arch, descending and/or thoraco-abdominal aortic aneurysm (N = 26 [47.3%]), dissection (N = 27 [49.1%], 6 acute/subacute), intramural haematoma (N = 1 [1.8%]), or aortic coarctation (N = 1 [1.8%)]. Device deployment was performed in aortic zone 0 (N = 16 [29.1%]), zone 1 (N = 5 [9.1%]), or zone 2 (N = 34 [61.8%]). Twenty-four (43.6%) underwent an aortic arch debranching procedure prior to TEVAR, and 23 (41.8%) had preoperative spinal drain placement. Technical success was 100.0% following the index operation. Operative mortality was 3.6% (N = 2). There were no cases of stroke or paralysis and one case of renal failure requiring dialysis. One patient experienced paraparesis treated with spinal drain placement with immediate resolution. Median follow-up was 366 (interquartile range [IQR]: 189, 444) days. The rate of endoleak over the follow-up period was 16.4% (N = 9). Three cases were treated medically with ongoing surveillance, and the remaining 6 underwent intervention: 5 endovascular, 1 open. The linearized all-cause mortality rate was 11.0% per patient-year.

Conclusions

Real-world use of the Gore TBE device demonstrates its safety and versatility in addressing various aortic pathologies and deployment zones. Further follow-up is necessary to assess the durability of repair, and novel investigational devices may improve outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Endovascular ProceduresBlood Vessel ProsthesisAortic Aneurysm, ThoracicBlood Vessel Prosthesis ImplantationAorta, ThoracicHumansMaleFemaleRetrospective StudiesAgedMiddle AgedAortic DissectionProsthesis DesignTreatment OutcomeAged, 80 and overPostoperative Complications

Stichwörter der Autor:innen

aortic arch aneurysmaortic arch dissectionbranched endoprosthesisthoracic endovascular aortic repair (TEVAR)

Autor:innen und Institutionen

  • Aorta Center, Heart, Vascular, and Thoracic Institute, Cleveland Clinic, Cleveland, OH 44195, United States.
  • Department of Thoracic and Cardiovascular Surgery, Heart, Vascular, and Thoracic Institute, Cleveland Clinic, Cleveland, OH 44195, United States.
  • Department of Vascular Surgery, Heart, Vascular, and Thoracic Institute, Cleveland Clinic, Cleveland, OH 44195, United States.
  • Department of Cardiology, Heart, Vascular, and Thoracic Institute, Cleveland Clinic, Cleveland, OH 44195, United States.

ORCID-Profile: Xiaoying Lou, Milind Desai, Eric E Roselli

Förderung

  • High-Risk Cardiovascular Research Philanthropic Fund

Zitieren

APA

Lou, X., Vargo, P. R., Caputo, F., Lyden, S., Kramer, B., Koprivanac, M., Kalahasti, V., Desai, M., Roselli, E. E., & Aorta Center Investigators (2026). Initial Post-Commercialization Experience Using a Thoracic Branch Endoprosthesis: Broad Application to Real-World Patients. European Journal of Cardio-Thoracic Surgery. https://doi.org/10.1093/ejcts/ezaf452

Vancouver

Lou X, Vargo PR, Caputo F, Lyden S, Kramer B, Koprivanac M, et al. Initial Post-Commercialization Experience Using a Thoracic Branch Endoprosthesis: Broad Application to Real-World Patients. European Journal of Cardio-Thoracic Surgery. 2026. doi:10.1093/ejcts/ezaf452

BibTeX
@article{Lou2026,
  author = {Lou, Xiaoying and Vargo, Patrick R and Caputo, Francis and Lyden, Sean and Kramer, Benjamin and Koprivanac, Marijan and Kalahasti, Vidyasagar and Desai, Milind and Roselli, Eric E and Aorta Center Investigators, Aorta Center Investigators},
  title = {Initial Post-Commercialization Experience Using a Thoracic Branch Endoprosthesis: Broad Application to Real-World Patients},
  journal = {European Journal of Cardio-Thoracic Surgery},
  year = {2026},
  doi = {10.1093/ejcts/ezaf452}
}
Kostenlos starten →