Comparison between invasive cardiac output and left ventricular assist device flow parameter
European Journal of Cardio-Thoracic Surgery · 2024
DOI: 10.1093/ejcts/ezae383 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift European Journal of Cardio-Thoracic Surgery, 2024. Verfasst von Amitai Segev und 13 weiteren Autor:innen. Beteiligte Einrichtung: The Leviev Heart Center, Sheba Medical Center, Ramat-Gan, Israel. Hauptthemen laut MEDLINE: Heart-Assist Devices, Cardiac Output, Thermodilution. Die Arbeit nennt 23 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVES To evaluate the correlation between left ventricular assist device flow parameter and invasive cardiac output measurements. METHODS We retrospectively evaluated right heart catheterization examinations performed in left ventricular assist device patients from 2 tertiary medical centres. We evaluated the correlation between cardiac output measurement methods (indirect Fick and thermodilution) and pump flow parameter using linear regression, and the agreement was graphically displayed using Bland–Altman plot technique. Clinical, echocardiographic, pump and haemodynamic parameters were compared between patients with and without discordance, defined as at least a 20% difference between measurements. RESULTS The study population consisted of 102 patients [median age 58 (51–64), 86% males, 17 ± 12 months post left ventricular assist device implantation] with a total of 544 measurements compared. Discordance between measurements was present in 102 of 226 (45%) comparisons between indirect Fick and pump flow and in 72 of 161 (48%) between thermodilution and pump flow. A comparison of indirect Fick and left ventricular assist device exhibited a statistical correlation of R = 0.751, and that of thermodilution and left ventricular assist device of R = 0.789. Parameters associated with the presence of discordance between cardiac output measurements included a higher rate of aortic valve opening, lower indirect Fick and higher thermodilution cardiac output. After excluding the lowest tertile of indirect Fick cardiac output values, the correlation between measurements improved (thermodilution: R = 0.879 and indirect Fick: R = 0.843, P < 0.001). CONCLUSIONS The current left ventricular assist device flow parameter provides an estimation of cardiac output that correlates well with indirect Fick and exhibits the strongest correlation with thermodilution. This correlation was stronger after excluding lower cardiac output values.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The Leviev Heart Center, Sheba Medical Center, Ramat-Gan, Israel.
- Clinic of Cardiac Transplantation and Mechanical Circulatory Support, Institute of Heart Diseases, University Hospital, Wroclaw Medical University, Wroclaw, Poland.
- Department of Physiology, Wroclaw Medical University, Wroclaw, Poland.
- Student Scientific Club of Transplantology and Advanced Therapies of Heart Failure, Wroclaw Medical University, Wrocław, Poland.
ORCID-Profile: Amitai Segev, Viana Copeland, Mateusz Sokolski, Elad Maor, Maksym Jura, Alexander Fardman, Avishay Grupper
Zitieren
APA
Segev, A., Copeland, V., Sokolski, M., Azaria, S., Morgan, A., Maor, E., Jura, M., Wilk, M., Przybylski, R., Freimark, D., Tal-Ben Ishay, R., Regev, U., Fardman, A., & Grupper, A. (2024). Comparison between invasive cardiac output and left ventricular assist device flow parameter. European Journal of Cardio-Thoracic Surgery. https://doi.org/10.1093/ejcts/ezae383
Vancouver
Segev A, Copeland V, Sokolski M, Azaria S, Morgan A, Maor E, et al. Comparison between invasive cardiac output and left ventricular assist device flow parameter. European Journal of Cardio-Thoracic Surgery. 2024. doi:10.1093/ejcts/ezae383
BibTeX
@article{Segev2024,
author = {Segev, Amitai and Copeland, Viana and Sokolski, Mateusz and Azaria, Sivan and Morgan, Avi and Maor, Elad and Jura, Maksym and Wilk, Mateusz and Przybylski, Roman and Freimark, Dov and Tal-Ben Ishay, Rotem and Regev, Udi and Fardman, Alexander and Grupper, Avishay},
title = {Comparison between invasive cardiac output and left ventricular assist device flow parameter},
journal = {European Journal of Cardio-Thoracic Surgery},
year = {2024},
doi = {10.1093/ejcts/ezae383}
}
EUVIMED