Mechanical failure analysis of patch materials used in aortic arch reconstruction: implications for clinical practice
European Journal of Cardio-Thoracic Surgery · 2023
DOI: 10.1093/ejcts/ezad366 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Cardio-Thoracic Surgery, 2023. Verfasst von Dominic P Recco und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac Surgery, Boston Children's Hospital, Boston, MA, USA. Hauptthemen laut MEDLINE: Aorta, Thoracic, Polyethylene Terephthalates. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NICHD NIH HHS, NIH.
Abstract (Originalsprache)
OBJECTIVES Thick-patch pulmonary homograft, autologous pericardium and CardioCel Neo are common patch materials for aortic arch reconstruction. Insufficient data exist on sutured patch strength and limits of use. We evaluated failure strength of these materials to develop a failure prediction model for clinical guidance. METHODS Patch failure strength was evaluated via sutured uniaxial and burst pressure testing. In sutured uniaxial testing, patches were sutured to aortic or Dacron tabs and pulled to failure. In burst pressure testing, patches were sewn into porcine aortas or Dacron grafts and pressurized to failure. Failure membrane tension was calculated. A prediction model of membrane tension versus vessel diameter was generated to guide clinical patch selection. RESULTS Combining sutured uniaxial and burst pressure test data, pulmonary homograft failure strength {0.61 [interquartile range (IQR): 0.44, 0.78] N/mm, n = 21} was less than half that of autologous pericardium [2.22 (IQR: 1.65, 2.78) N/mm, n = 15] and CardioCel Neo [1.31 (IQR: 1.20, 1.42) N/mm, n = 20]. Pulmonary homograft burst pressure [245 (IQR: 202, 343) mmHg, n = 7] was significantly lower than autologous pericardium [863 (IQR: 802, 919) mmHg, n = 6] and CardioCel Neo [766 (IQR: 721, 833) mmHg, n = 6]. Our model predicts failure limits for each patch material and outlines safety margins for combinations of aortic diameter and pressure. CONCLUSIONS Sutured failure strength of thick-patch pulmonary homograft was significantly lower than autologous pericardium and CardioCel Neo. Patient selection (predicted postoperative arch diameter and haemodynamics) and blood pressure management must be considered when choosing patch material for arch reconstruction. In older children and adolescents, autologous or bovine pericardium may be more suitable materials for aortic patch augmentation to minimize the risk of postoperative patch failure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Cardiac Surgery, Boston Children's Hospital, Boston, MA, USA.
- Harvard Medical School, Boston, MA, USA.
ORCID-Profile: Dominic P Recco, Peter E Hammer
Förderung
- NICHD NIH HHS
- NIH
Zitieren
APA
Recco, D. P., Kizilski, S. B., Marshall, L. E., Earley, P. D., Kneier, N. E., Del Nido, P. J., Hammer, P. E., & Hoganson, D. M. (2023). Mechanical failure analysis of patch materials used in aortic arch reconstruction: implications for clinical practice. European Journal of Cardio-Thoracic Surgery. https://doi.org/10.1093/ejcts/ezad366
Vancouver
Recco DP, Kizilski SB, Marshall LE, Earley PD, Kneier NE, Del Nido PJ, et al. Mechanical failure analysis of patch materials used in aortic arch reconstruction: implications for clinical practice. European Journal of Cardio-Thoracic Surgery. 2023. doi:10.1093/ejcts/ezad366
BibTeX
@article{Recco2023,
author = {Recco, Dominic P and Kizilski, Shannen B and Marshall, Lauren E and Earley, Patrick D and Kneier, Nicholas E and Del Nido, Pedro J and Hammer, Peter E and Hoganson, David M},
title = {Mechanical failure analysis of patch materials used in aortic arch reconstruction: implications for clinical practice},
journal = {European Journal of Cardio-Thoracic Surgery},
year = {2023},
doi = {10.1093/ejcts/ezad366}
}
EUVIMED