Aortic valve visualization and pressurization device: a novel device for intraoperative evaluation of aortic valve repair procedures

Bardia Arabkhani iD, Stefan C Sandker, Jerry Braun, Jesper Hjortnaes iD, Thomas J van Brakel, Dave R Koolbergen iD, Robert J M Klautz iD, Mark G Hazekamp

European Journal of Cardio-Thoracic Surgery · 2023

DOI: 10.1093/ejcts/ezad291 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Journal of Cardio-Thoracic Surgery, 2023. Verfasst von Bardia Arabkhani und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiothoracic surgery, Leiden University Medical Center (LUMC), Leiden, Netherlands. Hauptthemen laut MEDLINE: Cardiac Surgical Procedures, Aortic Valve Insufficiency. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES Aortic valve repair procedures are technically challenging, and current intraoperative evaluation methods often fail to predict the final echocardiographic result. We have developed a novel intraoperative aortic valve visualization and pressurization (AVP) device, enabling valve inspection under physiological conditions, and measuring aortic valve insufficiency (AI) during cardioplegic arrest. METHODS The AVP device is attached to the (neo)aorta, after any type of aortic valve repair, while the heart is arrested. The root is pressurized (60–80 mmHg) using a saline solution and an endoscope is introduced. The valve is inspected, and the amount of valvular leakage is measured. Postoperative ‘gold standard’ transesophageal echocardiogram measurements of AI are performed and compared against regurgitation volume measured. RESULTS In 24 patients undergoing valve-sparing root replacement, the AVP device was used. In 22 patients, postoperative echocardiographic AI was ≤ grade 1. The median leakage was 90 ml/min, IQR 60–120 ml/min. In 3 patients, additional adjustments after visual inspection was performed. In 2 patients, with complex anatomy, the valve was replaced. In one, after evaluation with the device, there was undesirable result visually and residual AI of 330 ml/min, and in another, 260 ml/min residual AI was measured and valve restriction on visual inspection. CONCLUSIONS The novel AVP device enables intraoperative evaluation of the valve under physiological conditions, while still on arrested heart, and allows for targeted adjustments. The AVP device can be an important aid for intraoperative evaluation of the aortic valve, during valve repair and valve-sparing procedures, thereby making the operative result more predictable and the operation more efficient.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac Surgical ProceduresAortic Valve InsufficiencyHumansAortic ValveAortaEchocardiographyTreatment Outcome

Stichwörter der Autor:innen

Aortic valve repairDeviceIntraoperativeValve-sparing root replacement

Autor:innen und Institutionen

  • Department of Cardiothoracic surgery, Leiden University Medical Center (LUMC), Leiden, Netherlands.
  • Department of Cardiothoracic surgery, Amsterdam UMC, Amsterdam, Netherlands.

ORCID-Profile: Bardia Arabkhani, Jesper Hjortnaes, Dave R Koolbergen, Robert J M Klautz

Zitieren

APA

Arabkhani, B., Sandker, S. C., Braun, J., Hjortnaes, J., van Brakel, T. J., Koolbergen, D. R., Klautz, R. J. M., & Hazekamp, M. G. (2023). Aortic valve visualization and pressurization device: a novel device for intraoperative evaluation of aortic valve repair procedures. European Journal of Cardio-Thoracic Surgery. https://doi.org/10.1093/ejcts/ezad291

Vancouver

Arabkhani B, Sandker SC, Braun J, Hjortnaes J, van Brakel TJ, Koolbergen DR, et al. Aortic valve visualization and pressurization device: a novel device for intraoperative evaluation of aortic valve repair procedures. European Journal of Cardio-Thoracic Surgery. 2023. doi:10.1093/ejcts/ezad291

BibTeX
@article{Arabkhani2023,
  author = {Arabkhani, Bardia and Sandker, Stefan C and Braun, Jerry and Hjortnaes, Jesper and van Brakel, Thomas J and Koolbergen, Dave R and Klautz, Robert J M and Hazekamp, Mark G},
  title = {Aortic valve visualization and pressurization device: a novel device for intraoperative evaluation of aortic valve repair procedures},
  journal = {European Journal of Cardio-Thoracic Surgery},
  year = {2023},
  doi = {10.1093/ejcts/ezad291}
}
Kostenlos starten →