Wideband black- and bright-blood late gadolinium enhancement imaging in patients with implantable cardiac devices at 1.5 T for improved visualization and localization of myocardial injury
European Heart Journal - Imaging Methods and Practice · 01.01.2026 · 4 (1)
DOI: 10.1093/ehjimp/qyag102 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Heart Journal - Imaging Methods and Practice, online erschienen am 30.05.2026. Verfasst von Pauline Gut und 11 weiteren Autor:innen. Beteiligte Einrichtung: IHU LIRYC, Heart Rhythm Disease Institute, Université de Bordeaux-INSERM U1045, Avenue du Haut Lévêque, Pessac 33604, France. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: French National Research Agency, French Federation of Cardiology, European Research Council.
Abstract (Originalsprache)
Aims
Late gadolinium enhancement (LGE) imaging in patients with implantable cardiac devices is often degraded by device-related artefacts. Wideband phase-sensitive inversion recovery (PSIR) mitigates these artefacts but suffers from low scar-blood contrast, limiting detection of subendocardial and small focal scars. Therefore, a novel wideband SPOT (Scar-specific imaging with Preserved myOcardial visualizaTion) sequence combining co-registered wideband black- and bright-blood LGE to improve scar-blood contrast and localization, with wideband inversion recovery and wideband T2 preparation for artefact reduction, was implemented at 1.5 T. Methods and results Wideband SPOT was compared with wideband PSIR, conventional PSIR, and conventional SPOT in an animal and 18 patients with devices and known myocardial lesions. Patient analyses included qualitative assessment of image quality, artefact severity, diagnostic confidence, and LGE segment detection; quantitative measurements of left ventricular (LV) volume, scar volume, and scar size; agreement analysis between wideband SPOT and wideband PSIR; and reproducibility analysis. In the animal, wideband SPOT reduced device-related artefacts and enabled accurate scar localization consistent with histology. In patients, wideband SPOT and wideband PSIR yielded similar image and artefact reduction (P > 0.05), but wideband SPOT reduced diagnostic uncertainty by 24% (P = 0.080). Inter-observer agreement for qualitative scores was moderate to good [intra-class correlation coefficient (ICC) 0.54–0.83, P < 0.001]. Quantitative LV and scar measurements demonstrated excellent correlation (r > 0.9, P < 0.01) and good-to-excellent reproducibility (ICC 0.85–0.98, P < 0.001) between techniques.
Conclusion
Compared with wideband PSIR, wideband SPOT showed a trend towards enhanced scar detection, identifying additional scarred segments while maintaining similar overall scar burden. It also improved inter-observer agreement, preserved anatomical reference, and reduced device artefacts, which may improve diagnostic confidence and risk stratification.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- IHU LIRYC, Heart Rhythm Disease Institute, Université de Bordeaux-INSERM U1045, Avenue du Haut Lévêque, Pessac 33604, France.
- Department of Diagnostic and Interventional Radiology, Lausanne University Hospital and University of Lausanne, Rue du Bugnon 46, Lausanne 1005, Switzerland.
- Department of Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, Avenue de Magellan, Pessac 33604, France.
- Heart & Vessels Department, Cardiology Division, Lausanne University Hospital, Rue du Bugnon 46, Lausanne 1005, Switzerland.
- Department of Cardiac Pacing and Electrophysiology, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, Avenue de Magellan, Pessac 33604, France.
- CIBM Center for Biomedical Imaging, EPFL AVP CP CIBM, Station 6, Lausanne 1015, Switzerland.
ORCID-Profile: Pauline Gut, Guido Caluori, Konstantinos Vlachos, Panagiotis Antiochos, Baptiste Durand, Ambra Masi, Juerg Schwitter, Frederic Sacher, Pierre Jaïs, Matthias Stuber, Aurélien Bustin
Förderung
- French National Research Agency
- French Federation of Cardiology
- European Research Council
- European Union
- Horizon 2020
Zitieren
APA
Gut, P., Cochet, H., Caluori, G., Vlachos, K., Antiochos, P., Durand, B., Masi, A., Schwitter, J., Sacher, F., Jaïs, P., Stuber, M., & Bustin, A. (2026). Wideband black- and bright-blood late gadolinium enhancement imaging in patients with implantable cardiac devices at 1.5 T for improved visualization and localization of myocardial injury. European Heart Journal - Imaging Methods and Practice, 4(1). https://doi.org/10.1093/ehjimp/qyag102
Vancouver
Gut P, Cochet H, Caluori G, Vlachos K, Antiochos P, Durand B, et al. Wideband black- and bright-blood late gadolinium enhancement imaging in patients with implantable cardiac devices at 1.5 T for improved visualization and localization of myocardial injury. European Heart Journal - Imaging Methods and Practice. 2026;4(1). doi:10.1093/ehjimp/qyag102
BibTeX
@article{Gut2026,
author = {Gut, Pauline and Cochet, Hubert and Caluori, Guido and Vlachos, Konstantinos and Antiochos, Panagiotis and Durand, Baptiste and Masi, Ambra and Schwitter, Juerg and Sacher, Frederic and Jaïs, Pierre and Stuber, Matthias and Bustin, Aurélien},
title = {Wideband black- and bright-blood late gadolinium enhancement imaging in patients with implantable cardiac devices at 1.5 T for improved visualization and localization of myocardial injury},
journal = {European Heart Journal - Imaging Methods and Practice},
year = {2026},
volume = {4},
number = {1},
doi = {10.1093/ehjimp/qyag102}
}