Influence of the battery level of a handheld X-ray device on the objective image quality of radiographs
Dentomaxillofacial Radiology · 2026
DOI: 10.1093/dmfr/twaf050 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2026. Verfasst von Laura Ricardina Ramirez-Sotelo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical-Surgical Stomatology, Faculty of Dentistry, National University of San Marcos, Lima 15081, Peru. Hauptthemen laut MEDLINE: Electric Power Supplies, Radiography, Dental, Radiographic Image Enhancement, Radiography, Dental, Digital. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 26 Literaturangaben. Gefördert von: Fundação de Amparo à Pesquisa do Estado de São Paulo, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior.
Abstract (Originalsprache)
Objectives
To evaluate the influence of the battery level of a handheld X-ray device on the objective image quality of radiographs.
Methods
An acrylic-block was radiographed using a SnapShot solid-state sensor and an Eagle handheld X-ray device set at 60 kVp and 0.88 mAs, with the battery at ten different levels (ranging from fully charged to 10%). Six radiographs were acquired at each level. Mean grey values and standard deviations were collected using ImageJ software to assess image brightness, noise, and uniformity. Additionally, an aluminium step-wedge was radiographed under the same conditions for contrast evaluation. Image quality parameters were compared across different battery levels using analysis of variance (ANOVA) with Tukey’s post-hoc test (α = 0.05).
Results
Images obtained with a 100% battery level showed higher brightness than those at 50% (P < 0.0001); the battery level did not affect image noise (P = 0.091); uniformity at 20% battery level differed from all other levels (P < 0.0001); and contrast was highest at 90% battery compared to all other levels (P < 0.0001). Overall, the values remained consistent across different battery levels.
Conclusions
Despite some significant differences, the objective image quality remained stable throughout the entire battery charge cycle of the tested handheld X-ray device.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical-Surgical Stomatology, Faculty of Dentistry, National University of San Marcos, Lima 15081, Peru.
- Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba, SP 13414-903, Brazil.
ORCID-Profile: Laura Ricardina Ramirez-Sotelo, Débora Costa Ruiz, Amanda Farias-Gomes, Matheus L Oliveira, Deborah Queiroz Freitas, Francisco Haiter-Neto
Förderung
- Fundação de Amparo à Pesquisa do Estado de São Paulo
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
Zitieren
APA
Ramirez-Sotelo, L. R., Ruiz, D. C., Farias-Gomes, A., Oliveira, M. L., Freitas, D. Q., & Haiter-Neto, F. (2026). Influence of the battery level of a handheld X-ray device on the objective image quality of radiographs. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twaf050
Vancouver
Ramirez-Sotelo LR, Ruiz DC, Farias-Gomes A, Oliveira ML, Freitas DQ, Haiter-Neto F. Influence of the battery level of a handheld X-ray device on the objective image quality of radiographs. Dentomaxillofacial Radiology. 2026. doi:10.1093/dmfr/twaf050
BibTeX
@article{RamirezSotelo2026,
author = {Ramirez-Sotelo, Laura Ricardina and Ruiz, Débora Costa and Farias-Gomes, Amanda and Oliveira, Matheus L and Freitas, Deborah Queiroz and Haiter-Neto, Francisco},
title = {Influence of the battery level of a handheld X-ray device on the objective image quality of radiographs},
journal = {Dentomaxillofacial Radiology},
year = {2026},
doi = {10.1093/dmfr/twaf050}
}
EUVIMED