Clinical application of a modified image receptor-holding device to reduce proximal surfaces overlapping in bitewing radiography

Lara Ravanelli iD, Michelle Chang iD, Henrique Mateus Alves Felizardo iD, Deivi Cascante-Sequeira iD, Hugo Gaêta-Araujo iD

Dentomaxillofacial Radiology · 2025

DOI: 10.1093/dmfr/twaf043 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2025. Verfasst von Lara Ravanelli und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Stomatology, Public Health, and Forensic Dentistry, Ribeirão Preto School of Dentistry, University of São Paulo (USP), Ribeirão Preto, …. Hauptthemen laut MEDLINE: Radiography, Bitewing, Radiography, Dental, Digital. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 8 Literaturangaben. Gefördert von: Fundação de Amparo à Pesquisa do Estado de São Paulo, Pró-Reitoria de Pesquisa e Inovação of the University of São Paulo, Programa de Apoio a Novos Docentes-Process n°.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To clinically validate the use of a modified image receptor-holding device (IRHD) for the bitewing technique.

Methods

Premolars and molars’ bitewing radiographs of 101 patients were acquired using 2 IRHD (traditional and modified). The modified IRHD was computer-aided designed and 3D printed. Images were acquired by a single operator using a digital radiographic system with phosphor storage plates. A questionnaire regarding the perception of the use of the 2 IRHD was applied to patients and the operator. Image evaluation (paired and individual) was performed by 2 oral and maxillofacial radiologists. Repetition, overlap rate, and operator/patient perception were expressed as absolute and relative frequencies, and compared between the devices using the chi-square test with a 5% significance level.

Results

The paired evaluation showed no difference in the rate of proximal surface overlap between the IRHDs (P ≥ .05). Individual evaluation showed a lower rate of proximal surface overlap with the modified IRHD, with emphasis on the distal surface in the premolar region (P < .05). The modified IRHD presented greater ease of positioning (52.3%), greater stability (64.9%), preference of use by the operator (54.5%), and greater comfort was reported by the patients (54.5%).

Conclusions

The modified IRHD may be more comfortable and has the potential to reduce the proximal surface overlap than the traditional IRHD, especially when there is an indication for a single radiograph in the premolar region. Advances in knowledge We clinically validated a modified IRHD that has potential to decrease proximal surfaces overlap with greater comfort for the patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiography, BitewingRadiography, Dental, DigitalHumansFemaleMaleAdultMiddle AgedPrinting, Three-DimensionalSurveys and QuestionnairesBicuspidEquipment DesignComputer-Aided DesignMolarAgedAdolescent

Stichwörter der Autor:innen

bitewing radiographydental cariesdental radiography

Autor:innen und Institutionen

  • Department of Stomatology, Public Health, and Forensic Dentistry, Ribeirão Preto School of Dentistry, University of São Paulo (USP), Ribeirão Preto, SP, 14040-904, Brazil.
  • Department of Dental Materials and Prosthesis, Ribeirão Preto School of Dentistry, University of São Paulo (USP), Ribeirão Preto, SP, 14040-904, Brazil.
  • Department of Diagnostic Sciences, University of Costa Rica School of Dentistry, San José, 11501-2060, Costa Rica.

ORCID-Profile: Lara Ravanelli, Michelle Chang, Henrique Mateus Alves Felizardo, Deivi Cascante-Sequeira, Hugo Gaêta-Araujo

Förderung

  • Fundação de Amparo à Pesquisa do Estado de São Paulo
  • Pró-Reitoria de Pesquisa e Inovação of the University of São Paulo
  • Programa de Apoio a Novos Docentes-Process n°
  • Programa de Apoio a Novos Docentes—Process n°

Zitieren

APA

Ravanelli, L., Chang, M., Felizardo, H. M. A., Cascante-Sequeira, D., & Gaêta-Araujo, H. (2025). Clinical application of a modified image receptor-holding device to reduce proximal surfaces overlapping in bitewing radiography. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twaf043

Vancouver

Ravanelli L, Chang M, Felizardo HMA, Cascante-Sequeira D, Gaêta-Araujo H. Clinical application of a modified image receptor-holding device to reduce proximal surfaces overlapping in bitewing radiography. Dentomaxillofacial Radiology. 2025. doi:10.1093/dmfr/twaf043

BibTeX
@article{Ravanelli2025,
  author = {Ravanelli, Lara and Chang, Michelle and Felizardo, Henrique Mateus Alves and Cascante-Sequeira, Deivi and Gaêta-Araujo, Hugo},
  title = {Clinical application of a modified image receptor-holding device to reduce proximal surfaces overlapping in bitewing radiography},
  journal = {Dentomaxillofacial Radiology},
  year = {2025},
  doi = {10.1093/dmfr/twaf043}
}
Kostenlos starten →