Interference of titanium and zirconia implants on dental-dedicated MR image quality: ex vivo and in vivo assessment

Katrine M Johannsen, Jennifer Christensen iD, Louise Hauge Matzen, Brian Hansen, Rubens Spin-Neto iD

Dentomaxillofacial Radiology · 01.02.2025 · 54 (2) · 132-139

DOI: 10.1093/dmfr/twae071 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, online erschienen am 18.12.2024. Verfasst von Katrine M Johannsen und 4 weiteren Autor:innen. Beteiligte Einrichtung: Section for Oral Radiology and Endodontics, Department of Dentistry and Oral Health, Aarhus University, 8000 Aarhus, Denmark. Hauptthemen laut MEDLINE: Zirconium, Titanium, Magnetic Resonance Imaging, Dental Implants. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 17 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To assess the impact of titanium and zirconia implants on dental-dedicated MR image (ddMRI) quality ex vivo (magnetic field distortion [MFD]) and in vivo (artefacts). Methods ddMR images were acquired (MAGNETOM Free.Max, 0.55 T, Siemens Healthineers AG, Forchheim, Germany) using a dental-dedicated coil (Rapid Biomedical, Rimpar, Germany). Ex vivo: three phantoms were manufactured: one agar-embedded titanium implant, one agar-embedded zirconia implant, and one control phantom (agar 1.5%). Field map analysis of images acquired at 0.55 T, 1.5 T, and 3.0 T (MAGNETOM Sola and MAGNETOM Lumina, respectively, Siemens Healthineers AG, Forchheim, Germany) was done to illustrate the extent and severity of MFD caused by the implants. In vivo (0.55 T only): a splint was designed to serve as an implant carrier, allowing diverse implant positions (0, 1, 2, or 5 implants). A volunteer was imaged using multiple pulse sequences. Three blinded observers scored the images twice for the presence, severity, and type of artefacts, illustrated by descriptive statistics and inter- and intra-observer reproducibility (kappa statistics).

Results

Ex vivo: titanium produced more severe MFD than zirconia. MFD extent and amplitude increased with field strength (0.55 T < 1.5 T < 3.0 T). In vivo: titanium produced more artefacts than zirconia, generally as signal voids in tooth crowns close to implants. Inter- and intra-observer reproducibility ranged from 0.28 to 0.64 and 0.32 to 0.57, respectively.

Conclusions

The prevalence of artefacts increased with magnetic field strength. Titanium generated larger MFD than zirconia. For both materials, artefacts were visible mainly in the crown area. Observer reproducibility needs improvement by dedicated ddMRI training.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ZirconiumTitaniumMagnetic Resonance ImagingDental ImplantsArtifactsDental MaterialsHumansPhantoms, Imaging

Stichwörter der Autor:innen

MRIartefactsfield strength

Substanzen

ZirconiumTitaniumzirconium oxideDental ImplantsDental Materials

Autor:innen und Institutionen

  • Section for Oral Radiology and Endodontics, Department of Dentistry and Oral Health, Aarhus University, 8000 Aarhus, Denmark.
  • Center of Functionally Integrative Neuroscience, Aarhus University, 8200 Aarhus N, Aarhus, Denmark.

ORCID-Profile: Jennifer Christensen, Rubens Spin-Neto

Zitieren

APA

Johannsen, K. M., Christensen, J., Matzen, L. H., Hansen, B., & Spin-Neto, R. (2025). Interference of titanium and zirconia implants on dental-dedicated MR image quality: ex vivo and in vivo assessment. Dentomaxillofacial Radiology, 54(2), 132-139. https://doi.org/10.1093/dmfr/twae071

Vancouver

Johannsen KM, Christensen J, Matzen LH, Hansen B, Spin-Neto R. Interference of titanium and zirconia implants on dental-dedicated MR image quality: ex vivo and in vivo assessment. Dentomaxillofacial Radiology. 2025;54(2):132-139. doi:10.1093/dmfr/twae071

BibTeX
@article{Johannsen2025,
  author = {Johannsen, Katrine M and Christensen, Jennifer and Matzen, Louise Hauge and Hansen, Brian and Spin-Neto, Rubens},
  title = {Interference of titanium and zirconia implants on dental-dedicated MR image quality: ex vivo and in vivo assessment},
  journal = {Dentomaxillofacial Radiology},
  year = {2025},
  volume = {54},
  number = {2},
  pages = {132--139},
  doi = {10.1093/dmfr/twae071}
}
Kostenlos starten →