Development and Validation of Facile RP-HPLC Method for Simultaneous Determination of Timolol Maleate, Moxifloxacin Hydrochloride, Diclofenac Sodium and Dexamethasone in Plasma, Aqueous Humor and Pharmaceutical Products
Journal of Chromatographic Science · 2023-08-19 · 61 (7) · 678-687
DOI: 10.1093/chromsci/bmac057 ↗At a glance
Journal article in Journal of Chromatographic Science, published online on 2022-07-24. Written by Aqeel Shahzad and 5 co-authors. Institution: Department of Pharmaceutical Analysis, School of Pharmacy, Xi'an Jiaotong University, Xi'an 710061, China. Major MEDLINE topics: Timolol, Aqueous Humor. Cited 3 times so far (Crossref); 24 references.
Abstract (original language)
The present study aimed to develop a validated RP-HPLC method for the simultaneous determination of timolol maleate (TM), moxifloxacin hydrochloride (MOXI), diclofenac sodium (DS) and dexamethasone (DEXA) in human plasma, bovine aqueous humor and pharmaceutical preparations. The chromatographic separation was studied using the C18 column. The chromatographic conditions, such as composition, pH, the flow rate of mobile phase, the temperature of column, wavelength of absorption and injection volume of the sample, were studied. The method was validated to confirm specificity, linearity and accuracy in accordance with an International Conference on Harmonization guidelines. The optimum conditions for separation included mobile phase 0.05 M monobasic phosphate buffer: acetonitrile (65:35 v/v), pH of buffer adjusted to 6.2 and the flow rate of 1 mL/minute. The optimum temperature of the column was found to be 35°C, absorption wavelength 265 nm and injection volume 50 μL. The baseline separation of all four drugs with good sensitivity, resolution, and a less than 15 min run time was achieved. The retention time of TM, MOXI, DS and DEXA were 4.3,5.7,9.9 and 13.5 minutes respectively. The limit of detection (LOD) values were 6.2, 4.8, 0.8 and 1.2 ng/mL for TM, MOXI, DS and DEXA, respectively, whereas their respective limit of quantification (LOQ) values was: were 42.6, 26.8, 5.6 and 6.2 ng/mL. The coefficient of variation for intra-day and inter-day were in the range of 0.32–1.57 and 1.29–3.07%, respectively. The method was found to be sensitive, precise and accurate in human plasma and bovine aqueous humor and can be applied for the quantification of these compounds in plasma, aqueous humor and pharmaceuticals.
Subjects
MeSH terms (MEDLINE)
Substances
Authors and institutions
- Department of Pharmaceutical Analysis, School of Pharmacy, Xi'an Jiaotong University, Xi'an 710061, China.
- Department of Pharmaceutics, Faculty of Pharmacy, Bahauddin Zakariya University, Multan, 60800, Pakistan.
- Institute of Microbiology, University of Agriculture Faisalabad, Punjab 38040, Pakistan.
- Faculty of Medicine and Health, University of New South Wales Sydney, NSW 2052, Australia.
ORCID profiles: Aqeel Shahzad, Qiang Fu
Cite
APA
Shahzad, A., Arshad, S., Zubair, F., Shahzad, S., Batool, F., & Fu, Q. (2023). Development and Validation of Facile RP-HPLC Method for Simultaneous Determination of Timolol Maleate, Moxifloxacin Hydrochloride, Diclofenac Sodium and Dexamethasone in Plasma, Aqueous Humor and Pharmaceutical Products. Journal of Chromatographic Science, 61(7), 678-687. https://doi.org/10.1093/chromsci/bmac057
Vancouver
Shahzad A, Arshad S, Zubair F, Shahzad S, Batool F, Fu Q. Development and Validation of Facile RP-HPLC Method for Simultaneous Determination of Timolol Maleate, Moxifloxacin Hydrochloride, Diclofenac Sodium and Dexamethasone in Plasma, Aqueous Humor and Pharmaceutical Products. Journal of Chromatographic Science. 2023;61(7):678-687. doi:10.1093/chromsci/bmac057
BibTeX
@article{Shahzad2023,
author = {Shahzad, Aqeel and Arshad, Sohail and Zubair, Faryal and Shahzad, Shakeel and Batool, Farzana and Fu, Qiang},
title = {Development and Validation of Facile RP-HPLC Method for Simultaneous Determination of Timolol Maleate, Moxifloxacin Hydrochloride, Diclofenac Sodium and Dexamethasone in Plasma, Aqueous Humor and Pharmaceutical Products},
journal = {Journal of Chromatographic Science},
year = {2023},
volume = {61},
number = {7},
pages = {678--687},
doi = {10.1093/chromsci/bmac057}
}
EUVIMED