1001 Intraocular Lens (IOL) Recalls: Identifying Root Causes and Lessons for Safer Cataract Surgery
British Journal of Surgery · 10.07.2026 · 113 (Supplement_6)
DOI: 10.1093/bjs/znag063.689 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Surgery, online erschienen am 10.07.2026. Verfasst von G Goh Khye Ler und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Aim
This systematic review aims to investigate causes of intraocular lens (IOL) recalls associated with cataract surgery in the UK, and to evaluate their impact on patients, and highlight strategies to strengthen manufacturing and regulatory practice.
Method
A systematic review was conducted following PRISMA guidelines. Searches were performed across MEDLINE, EMBASE, Web of Science, SCOPUS, and the UK Medicines and Healthcare products Regulatory Agency (MHRA) database using terms related to "intraocular lens", "recall", and "cataract surgery". Eligible studies and field safety notices were screened for details on recall cause, affected product, and, where available, patient outcomes. Data were narratively synthesised and categorised into manufacturing or post-implant reaction errors.
Results
Twenty IOL recalls were identified. The most common cause was manufacturing error (60%) - dominated by mislabelled lens diopters (n=8); followed by post-implant reactions due to contamination or material residues (n=8). Notable examples included aluminium contamination (HOYA 2013), phosphate cleaning residue-related calcification (Oculentis 2017), and polyvinylpyrrolidone obstruction elevating intraocular pressure (NIDEK 2023). Reported incidence of adverse effects ranged from 0.32-6.2%, largely manageable with medical or surgical intervention. Limited long-term follow-up and inconsistent reporting were found across all cases.
Conclusions
IOL recalls predominantly arise from preventable manufacturing and labelling errors. Improved equipment maintenance, residue-free processing, and stronger regulatory oversight are essential. Transparent post-market surveillance and unified reporting of patient outcomes will enhance safety standards and reduce recurrence of such adverse events in cataract surgery.
Zitieren
APA
Ler, G. G. K., & Azizi, S. (2026). 1001 Intraocular Lens (IOL) Recalls: Identifying Root Causes and Lessons for Safer Cataract Surgery. British Journal of Surgery, 113(Supplement_6). https://doi.org/10.1093/bjs/znag063.689
Vancouver
Ler GGK, Azizi S. 1001 Intraocular Lens (IOL) Recalls: Identifying Root Causes and Lessons for Safer Cataract Surgery. British Journal of Surgery. 2026;113(Supplement_6). doi:10.1093/bjs/znag063.689
BibTeX
@article{Ler2026,
author = {Ler, G Goh Khye and Azizi, S},
title = {1001 Intraocular Lens (IOL) Recalls: Identifying Root Causes and Lessons for Safer Cataract Surgery},
journal = {British Journal of Surgery},
year = {2026},
volume = {113},
number = {Supplement_6},
doi = {10.1093/bjs/znag063.689}
}