EUVIMEDEUVIMED entdecken
FallberichtOpen Access · CC BY 4.0

Post cataract surgery refractive surprise due to intraocular lens mislabeling

Saeed Shokoohi Rad, Mohammad Reza Ansari Astaneh, Mohammad Yaser Kiarudi, Seyed Hossein Ghavami Shahri, Hamid Reza Heidarzadeh iD

Clinical Case Reports · 2024-07 · 12 (7)

DOI: 10.1002/ccr3.9104 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Clinical Case Reports, online erschienen am 21.06.2024. Verfasst von Saeed Shokoohi Rad und 4 weiteren Autor:innen. Beteiligte Einrichtung: Eye Research Center Mashhad University of Medical Sciences Mashhad Iran. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Key Clinical Message If an intraocular lens (IOL) is incorrectly labeled, problems can arise for surgeons and patients. Precise biometry is important to accurately determine the IOL power and prevent the implantation of the wrong IOL. Labeling and verifying IOLs with accuracy is crucial to ensuring the best possible results of cataract surgery. Abstract Mislabeling of IOLs can cause unpredictable problems for surgeons. However, we can prevent incorrect IOLs by using precise biometry to determine accurate IOL power and safely implant the correct IOL. A 50‐year‐old female with no medical or ocular history came to our clinic complaining of decreased vision in both eyes that had been ongoing for several months. After being diagnosed with cataracts, primary angle closure suspect, and high hyperopia, the patient underwent phacoemulsification surgery. A posterior chamber IOL was implanted, and visco‐goniosynechialysis was performed. During follow‐up appointments, it was discovered that the patient had an uncorrected visual acuity of 20/50 in her right eye, which was corrected to 20/20 with a + 7.00 D lens. Upon further evaluation, it was determined that the source of the error was due to a manufacturing mislabeling of the IOL power. The patient then underwent successful IOL exchange surgery, and her best‐corrected visual acuity became 20/20 with no significant refraction. This case highlights an uncommon source of refractive surprise after phacoemulsification surgery, successfully managed with IOL exchange surgery using the same IOL power from a different brand.

Schlagwörter

Stichwörter der Autor:innen

cataractintraocular lensphacoemulsificationrefractive surprise

Autor:innen und Institutionen

  • Eye Research Center Mashhad University of Medical Sciences Mashhad Iran.

ORCID-Profile: Hamid Reza Heidarzadeh

Zitieren

APA

Shokoohi Rad, S., Ansari Astaneh, M. R., Kiarudi, M. Y., Ghavami Shahri, S. H., & Heidarzadeh, H. R. (2024). Post cataract surgery refractive surprise due to intraocular lens mislabeling. Clinical Case Reports, 12(7). https://doi.org/10.1002/ccr3.9104

Vancouver

Shokoohi Rad S, Ansari Astaneh MR, Kiarudi MY, Ghavami Shahri SH, Heidarzadeh HR. Post cataract surgery refractive surprise due to intraocular lens mislabeling. Clinical Case Reports. 2024;12(7). doi:10.1002/ccr3.9104

BibTeX
@article{ShokoohiRad2024,
  author = {Shokoohi Rad, Saeed and Ansari Astaneh, Mohammad Reza and Kiarudi, Mohammad Yaser and Ghavami Shahri, Seyed Hossein and Heidarzadeh, Hamid Reza},
  title = {Post cataract surgery refractive surprise due to intraocular lens mislabeling},
  journal = {Clinical Case Reports},
  year = {2024},
  volume = {12},
  number = {7},
  doi = {10.1002/ccr3.9104}
}
Kostenlos starten →