EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Proton pencil beam scanning lattice radiotherapy: technical implementation and comparison against photon-based delivery

Abdul S Khalid, Adam H Aitkenhead iD, Matthew Lowe iD, Matthew Clarke, John-William Warmenhoven iD, Francesca Fiorini iD, Thomas W S Satherley iD, Maxwell Robinson, Geoff S Higgins iD, Kristoffer Petersson iD

British Journal of Radiology · 01.09.2026 · 99 (1185) · 1806-1819

DOI: 10.1093/bjr/tqag125 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift British Journal of Radiology, online erschienen am 23.05.2026. Verfasst von Abdul S Khalid und 9 weiteren Autor:innen. Beteiligte Einrichtung: Christie Medical Physics and Engineering, The Christie NHS Foundation Trust, Manchester Academic Health Science Centre (MAHSC), Manchester M20 4BX, U…. Hauptthemen laut MEDLINE: Proton Therapy, Photons, Radiotherapy Planning, Computer-Assisted, Neoplasms. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Cancer Research UK, National Institute for Health and Care Research, Manchester Biomedical Research Center.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Spatially fractionated radiotherapy in the form of a lattice of high-dose regions within the tumor has potential to improve the therapeutic response compared to standard radiotherapy. We aimed (1) to develop and dosimetrically validate a proton pencil-beam scanning lattice radiotherapy (LRT) planning technique, (2) to compare against a photon VMAT LRT technique, and (3) to assess which clinical sites are feasible for treatment.

Methods

10 cases were selected, covering sites in the brain, neck, pelvis, abdomen and shoulder with gross tumor volumes greater than 100 cm3. LRT plans were developed for a Varian ProBeam system, aiming to deliver 20 Gy to the lattice vertices while minimizing dose to surrounding tissues. Plan-specific quality assurance (PSQA) was performed by independent Monte-Carlo calculation and physical dosimetric verification. End-to-end tests were performed using a CIRS head phantom containing EBT4 film. Plan quality was evaluated against VMAT LRT.

Results

Clinically acceptable proton LRT plans were produced, delivering 16-20 Gy to the lattice vertices, a mean PTV dose of 2 Gy and satisfying organ-at-risk dose constraints. PSQA and end-to-end tests met clinical tolerances for non-LRT plans. For large target volumes, proton LRT produced lattices with a greater number of vertices than VMAT LRT.

Conclusions

Proton LRT is technically viable using currently available delivery systems. Large tumors such as sarcoma are the most likely candidates for future clinical studies, but proton LRT is feasible for target volumes as small as 100 cm3. Advances in knowledge The study demonstrates the feasibility of proton LRT for early phase clinical trials.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyPhotonsRadiotherapy Planning, Computer-AssistedNeoplasmsRadiotherapy, Intensity-ModulatedHumansRadiotherapy DosagePhantoms, ImagingMonte Carlo MethodOrgans at Risk

Stichwörter der Autor:innen

SFRTVMATlattice radiotherapyproton therapy

Autor:innen und Institutionen

  • Christie Medical Physics and Engineering, The Christie NHS Foundation Trust, Manchester Academic Health Science Centre (MAHSC), Manchester M20 4BX, United Kingdom.
  • Division of Cancer Sciences, Faculty of Biology Medicine and Health, The University of Manchester, Manchester M13 9PL, United Kingdom.
  • Department of Radiotherapy Physics, Churchill Hospital, Oxford University Hospitals NHS Trust, Oxford OX3 7LE, United Kingdom.
  • Department of Oncology, University of Oxford, Oxford OX3 7DQ, United Kingdom.

ORCID-Profile: Adam H Aitkenhead, Matthew Lowe, John-William Warmenhoven, Francesca Fiorini, Thomas W S Satherley, Geoff S Higgins, Kristoffer Petersson

Förderung

  • Cancer Research UK
  • National Institute for Health and Care Research
  • Manchester Biomedical Research Center

Zitieren

APA

Khalid, A. S., Aitkenhead, A. H., Lowe, M., Clarke, M., Warmenhoven, J. W., Fiorini, F., Satherley, T. W. S., Robinson, M., Higgins, G. S., & Petersson, K. (2026). Proton pencil beam scanning lattice radiotherapy: technical implementation and comparison against photon-based delivery. British Journal of Radiology, 99(1185), 1806-1819. https://doi.org/10.1093/bjr/tqag125

Vancouver

Khalid AS, Aitkenhead AH, Lowe M, Clarke M, Warmenhoven JW, Fiorini F, et al. Proton pencil beam scanning lattice radiotherapy: technical implementation and comparison against photon-based delivery. British Journal of Radiology. 2026;99(1185):1806-1819. doi:10.1093/bjr/tqag125

BibTeX
@article{Khalid2026,
  author = {Khalid, Abdul S and Aitkenhead, Adam H and Lowe, Matthew and Clarke, Matthew and Warmenhoven, John-William and Fiorini, Francesca and Satherley, Thomas W S and Robinson, Maxwell and Higgins, Geoff S and Petersson, Kristoffer},
  title = {Proton pencil beam scanning lattice radiotherapy: technical implementation and comparison against photon-based delivery},
  journal = {British Journal of Radiology},
  year = {2026},
  volume = {99},
  number = {1185},
  pages = {1806--1819},
  doi = {10.1093/bjr/tqag125}
}
Kostenlos starten →