Quantification of myocardial iron and fat—an experimental study with photon-counting detector CT
British Journal of Radiology · 01.05.2026 · 99 (1181) · 866-871
DOI: 10.1093/bjr/tqag020 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Radiology, online erschienen am 25.01.2026. Verfasst von Philipp N Maintz und 8 weiteren Autor:innen. Beteiligte Einrichtung: Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, 8091 Zurich, Switzerland. Hauptthemen laut MEDLINE: Myocardium, Iron, Adipose Tissue, Tomography, X-Ray Computed. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
To determine the feasibility and accuracy of photon-counting detector (PCD)-CT for iron and fat quantification in the myocardium.
Methods
Cylindrical tubes were filled with porcine myocardium and iron citrate with iron concentrations of 0-20 mg Fe g−1. Dilution series were prepared with myocardium and iron (no-fat probes) and with 5% fat (fat probes). The tubes were positioned in a chest phantom and were scanned with a calcium-scoring protocol on a PCD-CT. A re-parameterized 3-material decomposition was used to separate iron and fat from myocardium.
Results
On virtual monoenergetic images, attenuation increased linearly with iron concentrations in both fat and no-fat probes. In no-fat probes, linear regression yielded a slope of 1.2 HU (mg Fe g−1)−1 with an intercept of 35.8 HU (R2 = 0.964). In the fat probes, the slope was similar at 1.1 HU (mg Fe g−1)−1, while the regression line shifted downwards by 6.1 HU with an intercept of 29.6 HU (R2 = 0.985). Iron maps separated fat from iron with calculated median fat fractions of 4.85 in the fat and 0.90 in the no-fat probes. In iron images, attenuation increased linearly with increasing iron concentrations, with similar slopes between fat and no-fat probes and negligible differences in the intercept.
Conclusions
Experimental evidence indicates the feasibility and accuracy of PCD-CT for iron and fat quantification in the myocardium. Iron-specific 3-material decomposition eliminates the confounding effect of fat on myocardial iron quantification. Advances in knowledge This study highlights the value of dual-energy CT with 3-material decomposition for quantifying iron and fat in the myocardium. Thus, CT could serve as alternative for the current reference standard MRI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, 8091 Zurich, Switzerland.
- Faculty of Medicine, Sigmund Freud University, 1020 Vienna, Austria.
- Department of Radiology and Nuclear Medicine, Maastricht University Medical Centre+, 6229 HX Maastricht, the Netherlands.
- KlusLab Research, 8032 Zurich, Switzerland.
ORCID-Profile: Thomas Flohr, Victor Mergen, Matthias Eberhard, Hatem Alkadhi
Zitieren
APA
Maintz, P. N., Demmert, T. T., Flohr, T., Klambauer, K., Moser, L. J., Mergen, V., Eberhard, M., Froehlich, J. M., & Alkadhi, H. (2026). Quantification of myocardial iron and fat—an experimental study with photon-counting detector CT. British Journal of Radiology, 99(1181), 866-871. https://doi.org/10.1093/bjr/tqag020
Vancouver
Maintz PN, Demmert TT, Flohr T, Klambauer K, Moser LJ, Mergen V, et al. Quantification of myocardial iron and fat—an experimental study with photon-counting detector CT. British Journal of Radiology. 2026;99(1181):866-871. doi:10.1093/bjr/tqag020
BibTeX
@article{Maintz2026,
author = {Maintz, Philipp N and Demmert, Tristan T and Flohr, Thomas and Klambauer, Konstantin and Moser, Lukas J and Mergen, Victor and Eberhard, Matthias and Froehlich, Johannes M and Alkadhi, Hatem},
title = {Quantification of myocardial iron and fat—an experimental study with photon-counting detector CT},
journal = {British Journal of Radiology},
year = {2026},
volume = {99},
number = {1181},
pages = {866--871},