Estimating brain and eye lens dose for the cardiologist in interventional cardiology—are the dose levels of concern?

Markus Hulthén iD, Virginia Tsapaki iD, Angeliki Karambatsakidou

British Journal of Radiology · 2024

DOI: 10.1093/bjr/tqae089 ↗

Kurz gefasst

Fachartikel in der Zeitschrift British Journal of Radiology, 2024. Verfasst von Markus Hulthén und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, 171 76 Stockholm, Sweden. Hauptthemen laut MEDLINE: Lens, Crystalline, Radiation Dosage, Phantoms, Imaging, Cardiologists. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 64 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To establish conversion coefficients (CCs), between mean absorbed dose to the brain and eye lens of the cardiologist and the air kerma-area product, PKA, for a set of projections in cardiac interventional procedures. Furthermore, by taking clinical data into account, a method to estimate the doses per procedure, or annual dose, is presented.

Methods

Thermoluminescence dosimeters were used together with anthropomorphic phantoms, simulating a cardiologist performing an interventional cardiac procedure, to estimate the CCs for the brain and eye lens dose for nine standard projections, and change in patient size and x-ray spectrum. Additionally, a single CC has been estimated, accounting for each projections fraction of use in the clinic and associated PKA using clinical data from the dose monitoring system in our hospital.

Results

The maximum CCs for the eye lens and segment of the brain, is 5.47 μGy/Gycm2 (left eye lens) and 1.71 μGy/Gycm2 (left brain segment). The corresponding weighted CCs: are 3.39 μGy/Gycm2 and 0.89 μGy/Gycm2, respectively.

Conclusions

Conversion coefficients have been established under actual scatter conditions, showing higher doses on the left side of the operator. Using modern interventional x-ray equipment, interventional cardiac procedures will not cause high radiation dose levels to the operator when a ceiling mounted shield is used, otherwise there is a risk that the threshold dose values for cataract will be reached. Advance in knowledge In addition to the CCs for the different projections, methods for deriving a single CC per cardiac interventional procedure and dose per year were introduced.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lens, CrystallineRadiation DosagePhantoms, ImagingCardiologistsBrainOccupational ExposureHumansThermoluminescent DosimetryCardiologyRadiography, Interventional

Stichwörter der Autor:innen

brain doseconversion coefficientseye lens doseradiation protectionradiation protection in interventional cardiology

Autor:innen und Institutionen

  • Department of Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, 171 76 Stockholm, Sweden.
  • Dosimetry and Medical Radiation Physics Section, Human Health Division, IAEA, Vienna, Austria.
  • Department of Oncology-Pathology, Karolinska Institutet, 171 76 Stockholm, Sweden.

ORCID-Profile: Markus Hulthén, Virginia Tsapaki

Zitieren

APA

Hulthén, M., Tsapaki, V., & Karambatsakidou, A. (2024). Estimating brain and eye lens dose for the cardiologist in interventional cardiology—are the dose levels of concern?. British Journal of Radiology. https://doi.org/10.1093/bjr/tqae089

Vancouver

Hulthén M, Tsapaki V, Karambatsakidou A. Estimating brain and eye lens dose for the cardiologist in interventional cardiology—are the dose levels of concern?. British Journal of Radiology. 2024. doi:10.1093/bjr/tqae089

BibTeX
@article{Hulthn2024,
  author = {Hulthén, Markus and Tsapaki, Virginia and Karambatsakidou, Angeliki},
  title = {Estimating brain and eye lens dose for the cardiologist in interventional cardiology—are the dose levels of concern?},
  journal = {British Journal of Radiology},
  year = {2024},
  doi = {10.1093/bjr/tqae089}
}
Kostenlos starten →