Improving image quality consistency and diagnostic accuracy in lower extremity CT angiography using a split-bolus contrast injection protocol

Qian Tian, Shumeng Zhu, Yannan Cheng, Jianying Li, Tingting Qu, Xiaoqian Jia, Le Cao, Lihong Chen, Jianxin Guo

British Journal of Radiology · 2024

DOI: 10.1093/bjr/tqae036 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift British Journal of Radiology, 2024. Verfasst von Qian Tian und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, China. Hauptthemen laut MEDLINE: Computed Tomography Angiography, Arteries. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Science Development Foundation of the First Affiliated Hospital of Xi'an Jiaotong University, Key R & D Plan Project of Shaanxi Province-Joint University, Science Development Foundation of the First Affiliated Hospital of Xi’an Jiaotong University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate the clinical value of using a split-bolus contrast injection protocol in improving image quality consistency and diagnostic accuracy in lower extremity CT angiography (CTA).

Methods

Fifty (mean age, 66 ± 12 years) and 39 (mean age, 66 ± 11 years) patients underwent CTA in the lower extremity arteries using split-bolus and fixed-bolus injection schemes, respectively. The objective and subjective image quality of the 2 groups were compared and the diagnostic efficacy for the degree of vessel stenosis was compared using digital subtraction angiography as the gold standard. A P < .05 was considered statistically significant.

Results

In comparison with the fixed-bolus scheme, the split-bolus scheme greatly improved the consistency of image quality of the low extremities by significantly increasing the arterial enhancement (337.87 ± 64.67HU vs. 254.74 ± 71.58HU, P < .001), signal-to-noise ratio (22.58 ± 11.64 vs. 7.14 ± 1.98, P < .001), and contrast-to-noise ratio (37.21 ± 10.46 vs. 31.10 ± 15.40, P = .041) in the infrapopliteal segment. The subjective image quality was better (P < .001) and the diagnostic accuracy was higher in the split-bolus group than in the fixed-bolus group (96.00% vs. 91.67%, P < .05, for diagnosing >50% stenosis, and 97.00% vs. 89.10%, P < .05, for diagnosing occlusion) for the infrapopliteal segment arteries.

Conclusions

Compared with the fixed-bolus injection scheme, the split-bolus injection scheme improves the image quality consistency and diagnostic accuracy especially for the infrapopliteal segment arteries in lower extremity CTA. Advances in knowledge (1) The split-bolus injection scheme of CTA of the lower extremity arteries improves the overall image quality, uniformity of contrast enhancement. (2) Compared with the fixed-bolus injection scheme, the split-bolus injection scheme especially improves the infrapopliteal segment arteries image quality and diagnostic efficacy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Computed Tomography AngiographyArteriesHumansMiddle AgedAgedConstriction, PathologicAngiography, Digital SubtractionLower ExtremityContrast Media

Stichwörter der Autor:innen

diagnostic efficacyimage qualitylower extremity CT angiographysplit-bolus injection scheme

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, China.
  • GE Healthcare, Computed Tomography Research Center, Beijing 100176, China.

Förderung

  • Science Development Foundation of the First Affiliated Hospital of Xi'an Jiaotong University
  • Key R & D Plan Project of Shaanxi Province-Joint University
  • Science Development Foundation of the First Affiliated Hospital of Xi’an Jiaotong University
  • Key R & D Plan Project of Shaanxi Province—Joint University

Zitieren

APA

Tian, Q., Zhu, S., Cheng, Y., Li, J., Qu, T., Jia, X., Cao, L., Chen, L., & Guo, J. (2024). Improving image quality consistency and diagnostic accuracy in lower extremity CT angiography using a split-bolus contrast injection protocol. British Journal of Radiology. https://doi.org/10.1093/bjr/tqae036

Vancouver

Tian Q, Zhu S, Cheng Y, Li J, Qu T, Jia X, et al. Improving image quality consistency and diagnostic accuracy in lower extremity CT angiography using a split-bolus contrast injection protocol. British Journal of Radiology. 2024. doi:10.1093/bjr/tqae036

BibTeX
@article{Tian2024,
  author = {Tian, Qian and Zhu, Shumeng and Cheng, Yannan and Li, Jianying and Qu, Tingting and Jia, Xiaoqian and Cao, Le and Chen, Lihong and Guo, Jianxin},
  title = {Improving image quality consistency and diagnostic accuracy in lower extremity CT angiography using a split-bolus contrast injection protocol},
  journal = {British Journal of Radiology},
  year = {2024},
  doi = {10.1093/bjr/tqae036}
}
Kostenlos starten →