Waveform data capture substantial variation in tidal volume and other respiratory parameters
Annals of the American Thoracic Society · 2026
DOI: 10.1093/annalsats/aaoag024 ↗Kurz gefasst
Fachartikel in der Zeitschrift Annals of the American Thoracic Society, 2026. Verfasst von Emily E Moin und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pulmonary, Allergy, and Critical Care, University of Pennsylvania, Philadelphia, PA, United States. Hauptthemen laut MEDLINE: Tidal Volume, Respiration, Artificial. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NHLBI NIH HHS, NIGMS NIH HHS, NIH HHS.
Abstract (Originalsprache)
Rationale
Patients receiving invasive mechanical ventilation (IMV) require accurate assessments of ventilator parameters. Documentation of these parameters in standard practice may fail to capture meaningful variation due to intermittent missingness.
Objectives
To assess variation in continuously-measured ventilator parameters and agreement with measurements documented as part of routine care in the electronic health record (EHR).
Methods
We performed a retrospective cohort study of patients receiving IMV in a medical intensive care unit from November 2024—March 2025. We compared the observed tidal volume, minute ventilation, peak inspiratory pressure, and positive end-expiratory pressure, measured continuously from device waveforms with intermittent EHR documentation. We calculated descriptive statistics and measures of agreement between these sources. Measurements and Main Results For 59 encounters, the median age was 65 (IQR, 59–72), 33 (56%) patients were male and 17 (29%) were Black. 34 (58%) patients died or were discharged to hospice. Among 358 patient-days of data, continuous measurements captured significantly more variation than EHR-documented measurements across all parameters. The largest errors were in observed tidal volume (mean absolute error 69 mL, 95% CI, 62–77 mL; correlation coefficient 0.540). Agreement in tidal volume was worse among patients receiving mandatory modes of ventilation (correlation coefficient 0.454).
Conclusions
Intermittent measurement of ventilator parameters fails to capture large variability observed in continuous, waveform-derived measurements. Poor agreement in parameters like tidal volume, even in mandatory modes of ventilation, highlights the potential for ventilator waveform data to improve care and advance research for patients with acute respiratory distress syndrome and others receiving IMV.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Pulmonary, Allergy, and Critical Care, University of Pennsylvania, Philadelphia, PA, United States.
- Palliative and Advanced Illness Research Center, University of Pennsylvania, Philadelphia, PA, United States.
- Penn-CHOP Lung Biology Institute, University of Pennsylvania, Philadelphia, PA, United States.
- Center for Clinical Epidemiology and Biostatistics, University of Pennsylvania, Philadelphia, PA, United States.
- Leonard Davis Institute of Health Economics, University of Pennsylvania, Philadelphia, PA, United States.
- Penn Medicine, Philadelphia, PA, United States.
ORCID-Profile: Emily E Moin
Förderung
- NHLBI NIH HHS
- NIGMS NIH HHS
- NIH HHS
Zitieren
APA
Moin, E. E., Bennett, R. M., Moffett, A. T., Schmid, B. E., Long, K., Seewald, N. J., Reilly, J. P., & Weissman, G. E. (2026). Waveform data capture substantial variation in tidal volume and other respiratory parameters. Annals of the American Thoracic Society. https://doi.org/10.1093/annalsats/aaoag024
Vancouver
Moin EE, Bennett RM, Moffett AT, Schmid BE, Long K, Seewald NJ, et al. Waveform data capture substantial variation in tidal volume and other respiratory parameters. Annals of the American Thoracic Society. 2026. doi:10.1093/annalsats/aaoag024
BibTeX
@article{Moin2026,
author = {Moin, Emily E and Bennett, Rachel M and Moffett, Alexander T and Schmid, Benjamin E and Long, Kevin and Seewald, Nicholas J and Reilly, John P and Weissman, Gary E},
title = {Waveform data capture substantial variation in tidal volume and other respiratory parameters},
journal = {Annals of the American Thoracic Society},
year = {2026},
doi = {10.1093/annalsats/aaoag024}
}