EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Comparison of Home and Office Blood Pressure Devices in the Clinical Setting

Stephen P Juraschek, Medha Vyavahare, Jennifer L Cluett, Ruth-Alma Turkson-Ocran, Kenneth J Mukamal iD, Anthony M Ishak

American Journal of Hypertension · 2024

DOI: 10.1093/ajh/hpad120 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Hypertension, 2024. Verfasst von Stephen P Juraschek und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Blood Pressure Determination, Hypertension. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NHLBI NIH HHS, NIH, NHLBI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND Self-measured blood pressure (SMBP) monitoring is increasingly used for remote hypertension management, but the real-world performance of home blood pressure (BP) devices is unknown. We examined BP measurements from patients’ home devices using the American Medical Association’s (AMA) SMBP Device Accuracy Test tool. METHODS Patients at a single internal medicine clinic underwent up to five seated, same-arm BP readings using a home device and an automated BP device (Omron HEM-907XL). Following the AMA’s three-step protocol, we used the patient’s home device for the first, second, and fourth measurements and the office device for the third and fifth (if needed) measurements. Device agreement failure was defined as an absolute difference in systolic BP >10 mm Hg between the home and office devices in either of two confirmatory steps. Performance was examined by brand (Omron vs. non-Omron). Moreover, we examined patient factors associated with agreement failure via logistic regression models adjusted for demographic characteristics. RESULTS We evaluated 152 patients (mean age 60 ± 15 years, 58% women, 31% Black) seen between October 2020 and November 2021. Device agreement failure occurred in 22.4% (95% CI: 16.4%, 29.7%) of devices tested, including 19.1% among Omron devices and 27.6% among non-Omron devices (P = 0.23). No patient characteristics were associated with agreement failure. CONCLUSIONS Over one-fifth of home devices did not agree based on the AMA SMBP device accuracy protocol. These findings confirm the importance of office-based device comparisons to ensure the accuracy of home BP monitoring.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Pressure DeterminationHypertensionHumansFemaleMiddle AgedAgedMaleBlood PressureReproducibility of ResultsSphygmomanometersBlood Pressure Monitoring, Ambulatory

Stichwörter der Autor:innen

blood pressurehome blood pressure monitoringhypertensionpharmacistself-measured blood pressurevalidation

Autor:innen und Institutionen

  • Department of Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts, USA.
  • Department of Medicine, Harvard Medical School, Boston, Massachusetts, USA.
  • Department of Medicine, Healthcare Associates, Beth Israel Deaconess Medical Center, Boston, Massachusetts, USA.

ORCID-Profile: Kenneth J Mukamal

Förderung

  • NHLBI NIH HHS
  • NIH
  • NHLBI
  • Healthcare Associates

Zitieren

APA

Juraschek, S. P., Vyavahare, M., Cluett, J. L., Turkson-Ocran, R. A., Mukamal, K. J., & Ishak, A. M. (2024). Comparison of Home and Office Blood Pressure Devices in the Clinical Setting. American Journal of Hypertension. https://doi.org/10.1093/ajh/hpad120

Vancouver

Juraschek SP, Vyavahare M, Cluett JL, Turkson-Ocran RA, Mukamal KJ, Ishak AM. Comparison of Home and Office Blood Pressure Devices in the Clinical Setting. American Journal of Hypertension. 2024. doi:10.1093/ajh/hpad120

BibTeX
@article{Juraschek2024,
  author = {Juraschek, Stephen P and Vyavahare, Medha and Cluett, Jennifer L and Turkson-Ocran, Ruth-Alma and Mukamal, Kenneth J and Ishak, Anthony M},
  title = {Comparison of Home and Office Blood Pressure Devices in the Clinical Setting},
  journal = {American Journal of Hypertension},
  year = {2024},
  doi = {10.1093/ajh/hpad120}
}
Kostenlos starten →