Bench Evaluation of the Sedaconda Inhaled Sedation Device: Hygrometric Properties and Device Resistance
Respiratory Care · 2025
DOI: 10.1089/respcare.12987 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Respiratory Care, 2025. Verfasst von François Lellouche und 2 weiteren Autor:innen. Beteiligte Einrichtung: Dr. Lellouche, Ms. Cyr, and Mr. Bouchard are affiliated with Centre de Recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec-U…. Hauptthemen laut MEDLINE: Respiration, Artificial, Procedural Sedation, Anesthetics, Inhalation. Die Arbeit nennt 61 Literaturangaben.
Abstract (Originalsprache)
Background
The Sedaconda Anesthetic Conserving Device (ACD-S) is used for inhaled sedation in intubated patients and has passive humidification properties linked to the carbon material. However, limited evaluations of the hygrometric performance are available. The main aim of the study was to assess the humidification performance of this device with different ventilator settings and ambient temperatures.
Methods
We assessed hygrometry of inspiratory gases and resistance of the Sedaconda ACD-S using a bench test. Hygrometric measurements were performed with 5 min ventilations from 4.8 to 18 L/min (at constant ambient temperature: 25 ± 1°C) and with 3 ambient temperatures, 22-24°C, 25 ± 1°C, and 28-30°C (at constant minute ventilation: 10 L/min). Simulation of physiologic expiratory gases with an absolute humidity of 35 mg H2O/L was used for main measurements. Some measurements were made with expiratory humidity of 27.5 mg H2O/L which corresponds to what is found in patients with moderate hypothermia. Three hygrometric measurements using the psychrometric method were performed for each tested condition. Device resistance was evaluated before utilization and after 96 h on the humidified bench and compared with a paper- and foam-based heat and moisture exchangers (HME) with similar dead spaces.
Results
In tested conditions, mean humidity output was 29.8 ± 1.3 mg H2O/L. We found slightly lower humidity output at low versus high minute ventilation (respectively 28.0 ± 0.3 vs 30.6 ± 0.3, P < .001) and lowest versus highest ambient temperature (29.6 ± 0.3 vs 31.0 ± 0.2 mg H2O/L, P = .02). These differences are probably not clinically important. We found a linear relationship between expiratory and inspiratory humidity (R2 = 0.98, P < .001). Resistance of the tested device was higher, in comparison with paper- and foam-based HMEs, and increased slightly after 96 h of use on the bench.
Conclusions
The Sedaconda ACD-S is a high-performance passive humidifier device, within all tested conditions of minute ventilation, tidal volume, and ambient temperature. Resistance was slightly higher compared with HMEs with similar dead space and different materials.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Dr. Lellouche, Ms. Cyr, and Mr. Bouchard are affiliated with Centre de Recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec-Université Laval, Québec, Canada.
Zitieren
APA
Lellouche, F., Cyr, L. A., & Bouchard, P. A. (2025). Bench Evaluation of the Sedaconda Inhaled Sedation Device: Hygrometric Properties and Device Resistance. Respiratory Care. https://doi.org/10.1089/respcare.12987
Vancouver
Lellouche F, Cyr LA, Bouchard PA. Bench Evaluation of the Sedaconda Inhaled Sedation Device: Hygrometric Properties and Device Resistance. Respiratory Care. 2025. doi:10.1089/respcare.12987
BibTeX
@article{Lellouche2025,
author = {Lellouche, François and Cyr, Laura-Ann and Bouchard, Pierre-Alexandre},
title = {Bench Evaluation of the Sedaconda Inhaled Sedation Device: Hygrometric Properties and Device Resistance},
journal = {Respiratory Care},
year = {2025},
doi = {10.1089/respcare.12987}
}
EUVIMED