The Influence of Pressurized Metered Dose Inhaler Actuator Orifice Diameter on Regional Extrathoracic Deposition of Inhaled Epinephrine

Conor A Ruzycki iD, Scott Tavernini, George Luciuk, Kevin W Stapleton, Warren H Finlay, Andrew R Martin

Journal of Aerosol Medicine and Pulmonary Drug Delivery · 01.05.2025 · 38 (3) · 106-117

DOI: 10.1089/jamp.2024.0039 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Aerosol Medicine and Pulmonary Drug Delivery, online erschienen am 26.05.2025. Verfasst von Conor A Ruzycki und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Faculty of Engineering, University of Alberta, Edmonton, Canada. Hauptthemen laut MEDLINE: Epinephrine, Metered Dose Inhalers, Bronchodilator Agents. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 46 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Extrathoracic deposition is a large source of in vivo variability in dosing for pressurized metered dose inhaler (pMDI) aerosols. A majority of previous studies have focused on only total extrathoracic deposition for pMDIs. The present work evaluates regional deposition within the extrathoracic region to better understand the impact of actuator orifice diameter and inhalation flow rate on extrathoracic deposition of a suspension pMDI formulation of epinephrine.

Methods

Regional deposition of a commercially available HFA (hydrofluoroalkane) suspension pMDI formulation of epinephrine was evaluated using plastic and metal versions of the newly developed sectioned Alberta Idealized Throat (s-AIT), divided into analogs of the oral cavity, the pharynx/larynx, and the upper trachea. Influences of actuator orifice diameter and inhaler insertion angle on regional extrathoracic deposition were evaluated in the plastic s-AIT at a 30 L/min flow rate, followed by additional testing in the metal s-AIT to evaluate effects across a range of flow rates (from 10 to 100 L/min).

Results

Actuator orifice was found to strongly influence regional extrathoracic deposition of a commercially available epinephrine HFA suspension pMDI aerosol, with smaller actuator orifices yielding reduced oral cavity deposition and increased distal-filter ( in vitro lung) deposition in both the plastic and metal s-AIT. Inhalation flow rate was found to strongly influence deposition in the metal s-AIT, with higher flow rates associated with reduced oral cavity deposition, increased pharynx/larynx deposition, and increased upper trachea deposition. Smaller orifices showed less variability in results as a function of inhaler insertion angle.

Conclusions

Actuator orifice diameter (spanning 0.22–0.42 mm) can strongly influence regional deposition of an HFA epinephrine suspension pMDI aerosol within the extrathoracic region. Smaller actuator orifices may provide reduced oral cavity deposition and increased delivery to the lungs. Smaller actuator orifices may also reduce variability in extrathoracic deposition that is associated with patient use aspects such as inhaler insertion angle.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

EpinephrineMetered Dose InhalersBronchodilator AgentsAdministration, InhalationEquipment DesignAerosolsHumansTracheaPharynxModels, AnatomicParticle Size

Stichwörter der Autor:innen

Alberta Idealized Throatextrathoracic depositioninhaled epinephrinepMDI actuator orifice

Substanzen

EpinephrineAerosolsBronchodilator Agents

Autor:innen und Institutionen

  • Department of Mechanical Engineering, Faculty of Engineering, University of Alberta, Edmonton, Canada.
  • Lovelace Biomedical, Albuquerque, New Mexico, USA.
  • Kokua Pharma Inc, Richmond, Canada.
  • Department of Biomedical Engineering, Faculty of Engineering, University of Alberta, Edmonton, Canada.

ORCID-Profile: Conor A Ruzycki

Zitieren

APA

Ruzycki, C. A., Tavernini, S., Luciuk, G., Stapleton, K. W., Finlay, W. H., & Martin, A. R. (2025). The Influence of Pressurized Metered Dose Inhaler Actuator Orifice Diameter on Regional Extrathoracic Deposition of Inhaled Epinephrine. Journal of Aerosol Medicine and Pulmonary Drug Delivery, 38(3), 106-117. https://doi.org/10.1089/jamp.2024.0039

Vancouver

Ruzycki CA, Tavernini S, Luciuk G, Stapleton KW, Finlay WH, Martin AR. The Influence of Pressurized Metered Dose Inhaler Actuator Orifice Diameter on Regional Extrathoracic Deposition of Inhaled Epinephrine. Journal of Aerosol Medicine and Pulmonary Drug Delivery. 2025;38(3):106-117. doi:10.1089/jamp.2024.0039

BibTeX
@article{Ruzycki2025,
  author = {Ruzycki, Conor A and Tavernini, Scott and Luciuk, George and Stapleton, Kevin W and Finlay, Warren H and Martin, Andrew R},
  title = {The Influence of Pressurized Metered Dose Inhaler Actuator Orifice Diameter on Regional Extrathoracic Deposition of Inhaled Epinephrine},
  journal = {Journal of Aerosol Medicine and Pulmonary Drug Delivery},
  year = {2025},
  volume = {38},
  number = {3},
  pages = {106--117},
  doi = {10.1089/jamp.2024.0039}
}
Kostenlos starten →