Accuracy of a Continuous Glucose Monitor During Pediatric Type 1 Diabetes Inpatient Admissions
Diabetes Technology & Therapeutics · 2024-02 · 26 (2) · 119-124
DOI: 10.1089/dia.2023.0375 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diabetes Technology & Therapeutics, online erschienen am 31.01.2024. Verfasst von Erin C Cobry und 8 weiteren Autor:innen. Beteiligte Einrichtung: Barbara Davis Center for Diabetes, University of Colorado Anschutz Medical Campus, Aurora, Colorado, USA. Hauptthemen laut MEDLINE: Diabetes Mellitus, Type 1. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIDDK NIH HHS.
Themen: Diabetes mellitus
Abstract (Originalsprache)
Objective
Continuous glucose monitors (CGMs) used for type 1 diabetes management are associated with lower hemoglobin A1c. CGMs are not approved for inpatient use, when close glucose monitoring and intensive insulin management are essential for optimal health. Accuracy data from adult hospitalizations have been published, but pediatric data are limited. Design and Methods: This retrospective review of Dexcom G6 data from youth with type 1 diabetes during hospitalization assessed CGMs and matched (within 5 min) point-of-care (POC) and laboratory glucose values. Glucose values >400 and 100 mg/dL and 15/20/30 mg/dL if value is ≤100 mg/dL.
Results
A total of 1120 POC and 288 laboratory-matched pairs were collected from 83 unique patients (median age 12.0 years, 68.7% non-Hispanic white, 54.2% male) during 100 admissions. For POC values, overall, MARD was 11.8%, that on the medical floor was 13.5%, and that in the intensive care unit was 7.9%. The MARD for all laboratory values was 6.5%. In total, 98% of matched pairs were within Clarke Error Grid A and B zones.
Conclusions
Findings from our pediatric population were similar to accuracy reported in hospitalized adults, indicating the potential role for CGM use during pediatric hospitalizations. Additional research is needed to assess accuracy under various conditions, including medication use, as well as development of safe hospital protocols for successful CGM implementation for routine inpatient care.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Barbara Davis Center for Diabetes, University of Colorado Anschutz Medical Campus, Aurora, Colorado, USA.
- Department of Biostatistics and Informatics, Colorado School of Public Health, Aurora, Colorado, USA.
ORCID-Profile: Erin C Cobry, Lauren A Waterman, Gregory P Forlenza, R Paul Wadwa
Förderung
- NIDDK NIH HHS
Zitieren
APA
Cobry, E. C., Pyle, L., Waterman, L. A., Forlenza, G. P., Towers, L., Karami, A. J., Jost, E., Berget, C., & Wadwa, R. P. (2024). Accuracy of a Continuous Glucose Monitor During Pediatric Type 1 Diabetes Inpatient Admissions. Diabetes Technology & Therapeutics, 26(2), 119-124. https://doi.org/10.1089/dia.2023.0375
Vancouver
Cobry EC, Pyle L, Waterman LA, Forlenza GP, Towers L, Karami AJ, et al. Accuracy of a Continuous Glucose Monitor During Pediatric Type 1 Diabetes Inpatient Admissions. Diabetes Technology & Therapeutics. 2024;26(2):119-124. doi:10.1089/dia.2023.0375
BibTeX
@article{Cobry2024,
author = {Cobry, Erin C and Pyle, Laura and Waterman, Lauren A and Forlenza, Gregory P and Towers, Lindsey and Karami, Angela J and Jost, Emily and Berget, Cari and Wadwa, R Paul},
title = {Accuracy of a Continuous Glucose Monitor During Pediatric Type 1 Diabetes Inpatient Admissions},
journal = {Diabetes Technology & Therapeutics},
year = {2024},
volume = {26},
number = {2},
pages = {119--124},
doi = {10.1089/dia.2023.0375}
}
EUVIMED