Next Generation Pharmaceutical Impactor (A New Impactor for Pharmaceutical Inhaler Testing). Part II: Archival Calibration

Virgil A Marple, Bernard A Olson, Kumaragovindham Santhanakrishnan, Jolyon P Mitchell, Sharon C Murray, Buffy L Hudson-Curtis

Journal of Aerosol Medicine · 2003-09 · 16 (3) · 301-324

DOI: 10.1089/089426803769017668 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Aerosol Medicine, 2003-09. Verfasst von Virgil A Marple und 5 weiteren Autor:innen. Beteiligte Einrichtung: Mechanical Engineering Department, University of Minnesota, Minneapolis, MN, USA. Hauptthemen laut MEDLINE: Aerosols, Calibration. Die Arbeit wurde bisher 132-mal zitiert (Crossref) und nennt 9 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A new seven-stage cascade impactor, the Next Generation Pharmaceutical Impactor (NGI), has been developed for the pharmaceutical industry. A calibration following "good laboratory practice (GLP)" procedures has been performed on a specific archival NGI, deemed to be representative of all NGIs. Thus, this impactor had nozzle dimensions for each stage manufactured close to the middle of the tolerance band for the design specification, and therefore the average nozzle diameter was equal to the nominal value for that stage. An essential feature of the NGI is that it is designed to operate at any flow rate between 30 and 100 L/min. Thus, the calibration was made at inlet flow rates of 30, 60 and 100 L/min representing the lower bound, mid-region and upper bound of the specified range of operation for the impactor. The calibration data were then used to develop equations that predict the particle cut size for all components of the impactor at any flow rate from 30 to 100 L/min.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AerosolsCalibrationAdministration, InhalationEquipment DesignHumansNebulizers and VaporizersParticle Size

Substanzen

Aerosols

Autor:innen und Institutionen

  • Mechanical Engineering Department, University of Minnesota, Minneapolis, MN, USA.

Zitieren

APA

Marple, V. A., Olson, B. A., Santhanakrishnan, K., Mitchell, J. P., Murray, S. C., & Hudson-Curtis, B. L. (2003). Next Generation Pharmaceutical Impactor (A New Impactor for Pharmaceutical Inhaler Testing). Part II: Archival Calibration. Journal of Aerosol Medicine, 16(3), 301-324. https://doi.org/10.1089/089426803769017668

Vancouver

Marple VA, Olson BA, Santhanakrishnan K, Mitchell JP, Murray SC, Hudson-Curtis BL. Next Generation Pharmaceutical Impactor (A New Impactor for Pharmaceutical Inhaler Testing). Part II: Archival Calibration. Journal of Aerosol Medicine. 2003;16(3):301-324. doi:10.1089/089426803769017668

BibTeX
@article{Marple2003,
  author = {Marple, Virgil A and Olson, Bernard A and Santhanakrishnan, Kumaragovindham and Mitchell, Jolyon P and Murray, Sharon C and Hudson-Curtis, Buffy L},
  title = {Next Generation Pharmaceutical Impactor (A New Impactor for Pharmaceutical Inhaler Testing). Part II: Archival Calibration},
  journal = {Journal of Aerosol Medicine},
  year = {2003},
  volume = {16},
  number = {3},
  pages = {301--324},
  doi = {10.1089/089426803769017668}
}