Assessment of experimental values of effective energies and beam quality correction factors for out-of-field dosimetry in external beam radiotherapy using radiophotoluminescent glass dosimeters

Meïssane M'Hamdi, Pauline Maury, Mesut Aydin, Marc André Boivin, Nathan Benzazon iD, Julie Colnot, Rodrigue S Allodji iD, Eric Deutsch, Charlotte Robert iD, Ibrahima Diallo

Biomedical Physics & Engineering Express · 01.08.2026 · 12 (4) · 045027

DOI: 10.1088/2057-1976/ae87ab ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 16.07.2026. Verfasst von Meïssane M'Hamdi und 9 weiteren Autor:innen. Beteiligte Einrichtung: U1355 Radiothérapie moléculaire et innovations thérapeutiques, Université Paris-Saclay Gustave Roussy INSERM, Villejuif, France. Hauptthemen laut MEDLINE: Radiation Dosimeters, Glass, Radiometry, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Euratom Research and Training Programme, Agence Nationale de la Recherche, Institut National de la Santé et de la Recherche Médicale.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In photon external beam radiotherapy, out-of-field doses resulting from photon scattering and leakage are unavoidable and may contribute to radiation-induced side effects. However, modern treatment planning systems are not commissioned to accurately estimate doses outside the treatment field. This underscores the need for dedicated experimental protocols for accurate measurements. Radiophotoluminescent glass dosimeters (RPLGDs) are well suited for this task due to their low detection threshold. Yet, because they are energy-dependent, beam quality calibration is required. The purpose was to experimentally measure effective energies of photons outside the treatment field and to define beam quality correction factors for the GD-302M and GD-352M RPLGDs to be used in out-of-field dosimetry. A solid phantom was assembled using water-equivalent plates. GD-302M and GD-352M RPLGDs were positioned at various distances from the beam central axis (0, 10, 20, 30, and 40 cm), at different depths (1, 10, and 15 cm), and measurements were performed for multiple field sizes (5, 10, and 15 cm). The experiment was repeated using a PTW30010 ionization chamber. Different beam qualities were tested, including 6 and 18 MV photon beams from an Elekta VERSA HD and 6 and 20 MV photon beams from a Varian Clinac 2300 CD. The low-energy filter in the RPL GD-352M modifies the signal relative to the unfiltered GD-302M. We exploit this differential to estimate out-of-field effective energy, using a calibration curve derived from literature data based on the GD-302M/GD-352M signal ratio. Across all beam qualities and configurations tested, effective energies ranged from 180 keV (±15 keV) to 1302 keV (±106 keV). Effective energy decreased as the distance from the axis increased, with an upward trend beyond 30 cm from the beam axis. Beam quality correction factors for GD-302M RPLGDs varied between 0.745 (±0.002) and 1.284 (±0.027), while for GD-352M they ranged from 0.568 (±0.004) to 1.649 (±0.123). Feature analysis showed that, for the beam quality correction factors, off-axis distance and depth were the most important variables, with importance values of 39.3% (±2.7) and 35.0% (±1.8) for the GD-302M, and 29.0% (±0.8) and 47.5% (±1.8) for the GD-352M, respectively. For the effective energies, the importance of off-axis distance and depth variables was 28.1% (±0.6) and 22.6% (±1.3), respectively. In this work, we determined experimentally effective photon energies outside the treatment field. We developed a methodology to establish out-of-field distance-dependent beam quality correction factors for RPLGDs. The results demonstrate the need to take into account those correction factors to ensure accurate out-of-field dose measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosimetersGlassRadiometryRadiotherapy Planning, Computer-AssistedPhotonsPhantoms, ImagingCalibrationRadiotherapy DosageHumansScattering, Radiation

Stichwörter der Autor:innen

energy correction factorsout-of-field dosimetryperipheral dosephoton radiotherapyradiophotoluminescent glass dosimeter

Autor:innen und Institutionen

  • U1355 Radiothérapie moléculaire et innovations thérapeutiques, Université Paris-Saclay Gustave Roussy INSERM, Villejuif, France.
  • Département de Radiothérapie, Institut Gustave Roussy, Villejuif, France.
  • U1018, Epidémiologie des radiations, INSERM, Gustave Roussy, Université Paris-Saclay, Villejuif, France.

ORCID-Profile: Nathan Benzazon, Rodrigue S Allodji, Charlotte Robert

Förderung

  • Euratom Research and Training Programme
  • Agence Nationale de la Recherche
  • Institut National de la Santé et de la Recherche Médicale

Zitieren

APA

M'Hamdi, M., Maury, P., Aydin, M., Boivin, M. A., Benzazon, N., Colnot, J., Allodji, R. S., Deutsch, E., Robert, C., & Diallo, I. (2026). Assessment of experimental values of effective energies and beam quality correction factors for out-of-field dosimetry in external beam radiotherapy using radiophotoluminescent glass dosimeters. Biomedical Physics & Engineering Express, 12(4), 045027. https://doi.org/10.1088/2057-1976/ae87ab

Vancouver

M'Hamdi M, Maury P, Aydin M, Boivin MA, Benzazon N, Colnot J, et al. Assessment of experimental values of effective energies and beam quality correction factors for out-of-field dosimetry in external beam radiotherapy using radiophotoluminescent glass dosimeters. Biomedical Physics & Engineering Express. 2026;12(4):045027. doi:10.1088/2057-1976/ae87ab

BibTeX
@article{MHamdi2026,
  author = {M'Hamdi, Meïssane and Maury, Pauline and Aydin, Mesut and Boivin, Marc André and Benzazon, Nathan and Colnot, Julie and Allodji, Rodrigue S and Deutsch, Eric and Robert, Charlotte and Diallo, Ibrahima},
  title = {Assessment of experimental values of effective energies and beam quality correction factors for out-of-field dosimetry in external beam radiotherapy using radiophotoluminescent glass dosimeters},
  journal = {Biomedical Physics & Engineering Express},
  year = {2026},
  volume = {12},
  number = {4},
  pages = {045027},
  doi = {10.1088/2057-1976/ae87ab}
}
Kostenlos starten →