Algorithm development for automatic laser alignment assessment on an ACR CT phantom and its evaluation on sixteen CT scanners
Biomedical Physics & Engineering Express · 2023
DOI: 10.1088/2057-1976/acff76 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, 2023. Verfasst von Choirul Anam und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, In…. Hauptthemen laut MEDLINE: Software, Algorithms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 35 Literaturangaben. Gefördert von: Diponegoro University.
Abstract (Originalsprache)
Purpose
The aim of this study is to develop software to automatically assess the laser alignment on the ACR CT phantom and evaluate its accuracy on sixteen CT scanners.
Methods
Software for an automated method of laser alignment assessment on the ACR CT phantom was developed. Laser alignment assessment was based on the positions of the ball-bearing markers at the edge of the ACR CT phantom. The automatic assessment was performed using several steps, including segmentation to acquire the coordinates of the ball-bearing markers and determination of the distances between lines connecting them with lines through the center of the image. A comparison of the results from the automatic method with those from the manual method was performed. The manual measurements were carried out using MicroDicom Viewer. A Mann–Whitney U test was performed to determine the statistical difference between both methods. The evaluation was performed on images of the ACR CT phantom scanned with 16 CT scanners from 5 different CT manufacturers.
Results
The results confirmed that our software successfully segments the ball-bearing markers and determines the laser alignment assessment on the ACR CT phantom. Evaluation of the algorithm with images from the 16 CT scanners revealed that the difference between the results from automatic and manual methods were about 0.2 mm with a p value of around 0.7 (no statistical difference). Misalignment in the y -axis was larger than the misalignment in the x-axisfor the majority of the scanners tested. It was found that the phantom tended to be placed 2 mm higher than the iso-center.
Conclusions
Software to automatically assess CT laser alignment with the ACR CT phantom was successfully developed and evaluated. The automatic assessment was comparable to manual assessment. In addition, the automatic method was user independent and fast.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia.
- Department of Applied Physics and Medical Imaging, California State University Channel Islands, Camarillo, CA 93012, United States of America.
ORCID-Profile: Choirul Anam, Ariij Naufal
Förderung
- Diponegoro University
Zitieren
APA
Anam, C., Amilia, R., Naufal, A., & Dougherty, G. (2023). Algorithm development for automatic laser alignment assessment on an ACR CT phantom and its evaluation on sixteen CT scanners. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/acff76
Vancouver
Anam C, Amilia R, Naufal A, Dougherty G. Algorithm development for automatic laser alignment assessment on an ACR CT phantom and its evaluation on sixteen CT scanners. Biomedical Physics & Engineering Express. 2023. doi:10.1088/2057-1976/acff76
BibTeX
@article{Anam2023,
author = {Anam, Choirul and Amilia, Riska and Naufal, Ariij and Dougherty, Geoff},
title = {Algorithm development for automatic laser alignment assessment on an ACR CT phantom and its evaluation on sixteen CT scanners},
journal = {Biomedical Physics & Engineering Express},
year = {2023},
doi = {10.1088/2057-1976/acff76}
}
EUVIMED