Self-lubricating anticoagulant surface (SLACS) mimicking heart valve tissue for improving biocompatibility of artificial blood pumps

Shiyu Cui, Yaoyao Chen iD, Jun Zhong, Xuefeng Zhang, Wei Wang, Yuhua Ren, Xiaodan Gou, Yang Cheng, Qing Yang, Feng Chen iD, Ying He iD, Lihong Fan iD

Biomedical Materials · 01.08.2026 · 21 (4) · 045002

DOI: 10.1088/1748-605x/ae7e2f ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Materials, online erschienen am 06.07.2026. Verfasst von Shiyu Cui und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061 Shaanxi, People's Republic of China. Hauptthemen laut MEDLINE: Extracorporeal Membrane Oxygenation, Anticoagulants, Biocompatible Materials, Heart Valves. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, the Shaanxi University joint project, the Scientific Research Development Fund of the First Affiliated Hospital of Xi’an Jiaotong University in China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Extracorporeal membrane oxygenation (ECMO) represents the most advanced modalit y of extracorporeal life support. The centrifugal pump head, a key component in direct contact with blood, demands excellent blood compatibility to ensure optimal performance. To address this requirement, a self-lubricating anticoagulant surface (SLACS) was fabricated on a roughened stainless steel substrate through physical adhesion and chemical bonding using a dimethylsiloxane gel (S-PDMS). In the blood slidin g experiment, blood remained on 316 l stainless steel for 6 s but slid off the SLACS in just 0.64 s, demonstrating exceptional resistance to blood adhesion and superior blood compatibility. Applying SLACS to the ECMO centrifugal pump head significantly enhanced its anticoagulant and antibacterial properties, while also exhibiting excellent safety, stability and durability. The feasibility of SLACS as a novel coating was validated through static tests, computational simulations and dynamic experiments, offering a promising solution for improving the safety of ECMO therapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Extracorporeal Membrane OxygenationAnticoagulantsBiocompatible MaterialsHeart ValvesCoated Materials, BiocompatibleHeart-Assist DevicesHeart Valve ProsthesisHumansAnimalsMaterials TestingSurface PropertiesStainless SteelDimethylpolysiloxanes

Stichwörter der Autor:innen

anticoagulantblood compatibilityextracorporeal membrane oxygenation (ECMO)self-lubricating

Substanzen

AnticoagulantsBiocompatible MaterialsStainless SteelCoated Materials, BiocompatibleDimethylpolysiloxanes

Autor:innen und Institutionen

  • Department of Cardiovascular Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061 Shaanxi, People's Republic of China.
  • School of Mechanical Engineering, Xi'an Jiaotong University, Xi'an 710049, People's Republic of China.
  • Key Laboratory of Education Ministry for Modern Design Rotor-Bearing System and the School of Mechanical Engineering, Xi'an Jiaotong University, Xi'an, 710049, People's Republic of China.
  • Department of Docimology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061 Shaanxi, People's Republic of China.
  • Clinical laboratory, Xi 'an Gaoling District Hospital, Xi'an, 710200 Shannxi, People's Republic of China.
  • State Key Laboratory for Manufacturing System Engineering and Shaanxi Key Laboratory of Photonics Technology for Information, School of Electronic Science and Engineering, Xi'an Jiaotong University, Xi'an 710049, People…

ORCID-Profile: Yaoyao Chen, Feng Chen, Ying He, Lihong Fan

Förderung

  • National Natural Science Foundation of China
  • the Shaanxi University joint project
  • the Scientific Research Development Fund of the First Affiliated Hospital of Xi’an Jiaotong University in China

Zitieren

APA

Cui, S., Chen, Y., Zhong, J., Zhang, X., Wang, W., Ren, Y., Gou, X., Cheng, Y., Yang, Q., Chen, F., He, Y., & Fan, L. (2026). Self-lubricating anticoagulant surface (SLACS) mimicking heart valve tissue for improving biocompatibility of artificial blood pumps. Biomedical Materials, 21(4), 045002. https://doi.org/10.1088/1748-605x/ae7e2f

Vancouver

Cui S, Chen Y, Zhong J, Zhang X, Wang W, Ren Y, et al. Self-lubricating anticoagulant surface (SLACS) mimicking heart valve tissue for improving biocompatibility of artificial blood pumps. Biomedical Materials. 2026;21(4):045002. doi:10.1088/1748-605x/ae7e2f

BibTeX
@article{Cui2026,
  author = {Cui, Shiyu and Chen, Yaoyao and Zhong, Jun and Zhang, Xuefeng and Wang, Wei and Ren, Yuhua and Gou, Xiaodan and Cheng, Yang and Yang, Qing and Chen, Feng and He, Ying and Fan, Lihong},
  title = {Self-lubricating anticoagulant surface (SLACS) mimicking heart valve tissue for improving biocompatibility of artificial blood pumps},
  journal = {Biomedical Materials},
  year = {2026},
  volume = {21},
  number = {4},
  pages = {045002},
  doi = {10.1088/1748-605x/ae7e2f}
}
Kostenlos starten →