Fachartikel in der Zeitschrift Journal of Neural Engineering, 2025. Verfasst von Christopher K Nguyen und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bioengineering, The University of Texas at Dallas, Richardson, TX, United States of America. Hauptthemen laut MEDLINE: Silicon Compounds, Carbon Compounds, Inorganic, Electrodes, Implanted. Die Arbeit nennt 56 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIDCD NIH HHS, National Institute on Deafness and Other Communication Disorders.
Abstract (Originalsprache)
Objective . Chronically implanted microelectrode arrays (MEAs) are used for stimulating and recording neural activity in research and clinical settings. However, their reliability can be compromised by insufficient encapsulation stability. Amorphous silicon carbide (a-SiC), a chemically stable and biocompatible material, has emerged as a potential thin-film encapsulation for MEAs. We aimed to evaluate thin-film a-SiC encapsulation using electrical-accelerated aging (EAA) and to demonstrate a methodology for obtaining acceleration factors for EAA by Weibull analysis. Approach . Interdigitated electrodes (IDEs) encapsulated with a-SiC were subjected to voltage cycling and stepped-voltage protocols to measure leakage currents in buffered saline at 37 °C. EAA employed incrementally increasing voltage biases over time to induce degradation and reveal failure mechanisms. Main results . IDEs exhibited a significant change in electrical behavior on exposure to saline, with failure initiating at specific voltages and accompanied by gas evolution at defect sites. Incremental voltage biasing revealed a capacitive-to-faradaic transition in leakage current response that was used as a failure criterion. Significance . Acceleration factors for voltage-driven accelerated aging of a-SiC thin-film encapsulation can be obtained by Weibull analysis using a mechanistic failure criterion. Breakdown occurs at processing-related defects in the a-SiC. This study demonstrates the use of EAA for evaluating failure in a-SiC thin-film encapsulation used in implantable MEAs. EAA is broadly applicable to thin-film MEAs and provides a highly relevant method of predicting implanted lifetimes of bioelectronics.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Bioengineering, The University of Texas at Dallas, Richardson, TX, United States of America.
- Department of Materials Science and Engineering, The University of Texas at Dallas, Richardson, TX, United States of America.
- Office of Research and Innovation, The University of Texas at Dallas, Richardson, TX, United States of America.
- Blackrock Neurotech, Salt Lake City, UT, United States of America.
ORCID-Profile: Christopher K Nguyen, Negar Geramifard, Yupeng Wu, Alexandra Joshi-Imre, Stuart F Cogan
Förderung
- NIDCD NIH HHS
- National Institute on Deafness and Other Communication Disorders
Zitieren
APA
Nguyen, C. K., Geramifard, N., Wu, Y., Bhatt, M., Joshi-Imre, A., Negi, S., & Cogan, S. F. (2025). Electrical characterization and accelerated aging of amorphous silicon carbide implantable encapsulation. Journal of Neural Engineering. https://doi.org/10.1088/1741-2552/ae2956
Vancouver
Nguyen CK, Geramifard N, Wu Y, Bhatt M, Joshi-Imre A, Negi S, et al. Electrical characterization and accelerated aging of amorphous silicon carbide implantable encapsulation. Journal of Neural Engineering. 2025. doi:10.1088/1741-2552/ae2956
BibTeX
@article{Nguyen2025,
author = {Nguyen, Christopher K and Geramifard, Negar and Wu, Yupeng and Bhatt, Madhav and Joshi-Imre, Alexandra and Negi, Sandeep and Cogan, Stuart F},
title = {Electrical characterization and accelerated aging of amorphous silicon carbide implantable encapsulation},
journal = {Journal of Neural Engineering},
year = {2025},
doi = {10.1088/1741-2552/ae2956}
}
EUVIMED