EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A miniaturized and low-energy subcutaneous optical telemetry module for neurotechnology

Zhangyu Xu, Nhan Duy Truong, Armin Nikpour, Omid Kavehei iD

Journal of Neural Engineering · 2023

DOI: 10.1088/1741-2552/acd147 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Neural Engineering, 2023. Verfasst von Zhangyu Xu und 3 weiteren Autor:innen. Beteiligte Einrichtung: University of Sydney, School of Biomed Engineering, Sydney, NSW 2006, Australia. Hauptthemen laut MEDLINE: Telemetry, Amplifiers, Electronic. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Australian Research Council, University of Sydney.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study presents a proof-of-concept optical telemetry module that leverages a single light-emitting diode (LED) to transmit data at a high bit rate while consuming low power and occupying a small area. Our experiments showed that we could achieve 108 Mbit s −1 and 54 Mbit s −1 back telemetry data rates for tissue thicknesses of 3 mm and 8 mm, respectively. Approach. The proposed module is designed to be powered by near-field coupling and achieve bidirectional communication by low-speed downlink from near-field communication. It aims to minimize the size of the implant while providing reliable transmission that meets the requirements of high-speed wireless communication from a multi-electrode array neurotechnology implant outside the body.

Results

The power consumption of the module is 1.57 mW, including the power consumption of related circuits, resulting in an efficiency of 14.5 pJ bit −1 , at a tissue thickness of 3 mm and a data rate of 108 Mbit. The use of an optical lens, combined with tissue scattering effect and optimized emission angle, makes the module robust to misalignments of up to ±5 mm and ±15° between the implantable and external units. The LED in the implantable unit is only 0.98 × 0.98 × 0.6 mm 3 , and the testing module is composed of discrete components and laboratory instruments. Significance. This work aims to show how it is possible to strike a balance between a small, reliable, and high-bit-rate data uplink between a neural implant and its proximal, wirelessly connected external unit. This optical telemetry module has the potential to be integrated into a significantly miniaturized system through an application-specific integrated circuit and can support up to 1000 channels of neural recordings, each sampled at 9 kSps with a 12-bit readout resolution.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TelemetryAmplifiers, ElectronicEquipment DesignElectrodes, ImplantedWireless Technology

Stichwörter der Autor:innen

low-energyneurotechnologyoptical-telemetry

Autor:innen und Institutionen

  • University of Sydney, School of Biomed Engineering, Sydney, NSW 2006, Australia.
  • BrainConnect Pty Ltd, Sydney, NSW 2006, Australia.
  • Univ Sydney, Fac Med & Hlth, Cent Clin Sch, Sydney, NSW 2006, Australia.
  • Royal Prince Alfred Hosp, Dept Neurol, Camperdown, NSW 2050, Australia.

ORCID-Profile: Omid Kavehei

Förderung

  • Australian Research Council
  • University of Sydney

Zitieren

APA

Xu, Z., Truong, N. D., Nikpour, A., & Kavehei, O. (2023). A miniaturized and low-energy subcutaneous optical telemetry module for neurotechnology. Journal of Neural Engineering. https://doi.org/10.1088/1741-2552/acd147

Vancouver

Xu Z, Truong ND, Nikpour A, Kavehei O. A miniaturized and low-energy subcutaneous optical telemetry module for neurotechnology. Journal of Neural Engineering. 2023. doi:10.1088/1741-2552/acd147

BibTeX
@article{Xu2023,
  author = {Xu, Zhangyu and Truong, Nhan Duy and Nikpour, Armin and Kavehei, Omid},
  title = {A miniaturized and low-energy subcutaneous optical telemetry module for neurotechnology},
  journal = {Journal of Neural Engineering},
  year = {2023},
  doi = {10.1088/1741-2552/acd147}
}
Kostenlos starten →