Assessment of the calibration capability of KRISS for noble-metal thermocouples
Metrologia · 18.11.2026
DOI: 10.1088/1681-7575/aec3aa ↗Kurz gefasst
Fachartikel in der Zeitschrift Metrologia, online erschienen am 18.11.2026. Verfasst von Yong-Gyoo Kim und 3 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The Korea Research Institute of Standards and Science (KRISS) maintains the national temperature standards and provides high-accuracy calibration services exclusively for noble-metal thermocouples (Types R, S and B). Most industrial users prefer comparison calibration to primary fixed-point calibration because the latter is time-consuming and costly. Comparison calibrations, however, carry larger uncertainties, and values above 1100 °C must be obtained by extrapolation, which further increases the uncertainty. To establish the reliability of this practical approach, a comprehensive assessment of KRISS’s thermocouple calibration capability was undertaken.
Three Type R reference thermocouples were prepared, characterized for thermoelectric inhomogeneity, and calibrated at the ITS-90 fixed points (Sn, Zn, Al, Ag, Cu), and Pd melting point realized by the KRISS cell together with selected metal–carbon eutectic points (Fe–C, Co–C and Pd–C) up to 1560 °C. The same thermocouples were then mutually intercompared in a fully automated comparison system up to 1100 °C, after which the deviation functions were extrapolated to 1600 °C.
Temperature differences between the extrapolated comparison results and the independent fixed-point calibrations remained within ±0.3 °C across the entire range—substantially smaller than the expanded uncertainties assigned to routine comparison calibrations (0.8 °C from the ice point to 1100 °C and 1.6 °C from 1100 °C to 1600 °C). These results demonstrate that fixed-point-calibrated reference thermocouples can be used for routine comparison calibrations limited to 1100 °C with subsequent extrapolation to 1600 °C, while remaining fully consistent with the primary temperature scale. The combination of primary fixed-point realization, mutual inter-comparison, and validated extrapolation therefore provides a practical and fully traceable foundation for KRISS’s thermocouple calibration services.
Zitieren
APA
Kim, Y., Lee, Y. H., Kim, S., & Kwon, S. (2026). Assessment of the calibration capability of KRISS for noble-metal thermocouples. Metrologia. https://doi.org/10.1088/1681-7575/aec3aa
Vancouver
Kim Y, Lee YH, Kim S, Kwon S. Assessment of the calibration capability of KRISS for noble-metal thermocouples. Metrologia. 2026. doi:10.1088/1681-7575/aec3aa
BibTeX
@article{Kim2026,
author = {Kim, Yong-Gyoo and Lee, Young Hee and Kim, Sunghun and Kwon, Suyong},
title = {Assessment of the calibration capability of KRISS for noble-metal thermocouples},
journal = {Metrologia},
year = {2026},
doi = {10.1088/1681-7575/aec3aa}
}
EUVIMED