Design and optimization of a multi-module Compton camera system for thyroid scintigraphy
Physica Scripta · 22.05.2026 · 101 (20) · 205003
DOI: 10.1088/1402-4896/ae6a2d ↗Kurz gefasst
Fachartikel in der Zeitschrift Physica Scripta, online erschienen am 19.05.2026. Verfasst von Asuman Kolbaşı und 2 weiteren Autor:innen. Die Arbeit nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Directorate of Presidential Strategy and Budget of Türkiye, Scientific and Technological Research Council of Türkiye, Health Institutes of Türkiye.
Abstract (Originalsprache)
Thyroid scintigraphy is a common clinical nuclear medicine procedure that provides both metabolic and structural information about the thyroid gland. Current clinical practice employs pinhole or parallel-hole collimators mounted on gamma cameras, which present several limitations including significant weight, potential safety hazards, and reduced photon detection efficiency. This work presents the initial simulation results towards a novel compact Compton camera system specifically designed for thyroid scintigraphy applications that eliminates the need for mechanical collimation. The proposed system utilizes a dual-layer detector configuration with GAGG:Ce scintillator crystals: a 2 mm thick scattering layer and a 10 mm thick absorbing layer. Monte Carlo simulations using GAMOS were employed to optimize the detector geometry and validate imaging performance. Two reconstruction algorithms were implemented and compared: Compton cone back-projection and list-mode maximum likelihood expectation maximization (LM-MLEM). MLEM reconstruction with 8 iterations achieves around 4 mm spatial resolution at a 5 cm source distance and can detect isolated sub-millimeter features. To acquire standard thyroid views (LAO, anterior, RAO) simultaneously and resolve depth ambiguity present in single-module Compton cameras, a three-module setup was tested, showing significant improvement in 3D reconstruction. This configuration could shorten examination time and eliminate the need for patient repositioning required by conventional methods. The simulation results provide validated design parameters for experimental prototype construction currently underway in our laboratory, where the predicted performance will be verified through experimental measurements.
Förderung
- Directorate of Presidential Strategy and Budget of Türkiye
- Scientific and Technological Research Council of Türkiye
- Health Institutes of Türkiye
Zitieren
APA
Kolbaşı, A., Öztürk, C., & Akgün, B. (2026). Design and optimization of a multi-module Compton camera system for thyroid scintigraphy. Physica Scripta, 101(20), 205003. https://doi.org/10.1088/1402-4896/ae6a2d
Vancouver
Kolbaşı A, Öztürk C, Akgün B. Design and optimization of a multi-module Compton camera system for thyroid scintigraphy. Physica Scripta. 2026;101(20):205003. doi:10.1088/1402-4896/ae6a2d
BibTeX
@article{Kolba2026,
author = {Kolbaşı, Asuman and Öztürk, Cengizhan and Akgün, Bora},
title = {Design and optimization of a multi-module Compton camera system for thyroid scintigraphy},
journal = {Physica Scripta},
year = {2026},
volume = {101},
number = {20},
pages = {205003},
doi = {10.1088/1402-4896/ae6a2d}
}
EUVIMED