EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Accuracy of Skin Temperature Sensors Under Cold and Windy Conditions: Effects of Sensor Type and Insulation

Lisa Klous iD, Pim Scheiberlich iD, Aukje de Vrijer iD, Minke Pien Jolien Geerts iD, Anne Lamers iD, Koen Van Der Sanden, Boris Kingma

Physiological Measurement · 12.11.2026

DOI: 10.1088/1361-6579/aec1b2 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physiological Measurement, online erschienen am 12.11.2026. Verfasst von Lisa Klous und 6 weiteren Autor:innen. Beteiligte Einrichtung: Human Performance, TNO, Kampweg 55, 3769DE Soesterberg, Soesterberg, 3769DE, Netherlands. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dutch Ministry of Defense.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Accurate skin temperature measurement is essential for assessing thermal strain in cold environments. This study evaluated the accuracy of four contact sensors under cold and windy conditions and examined how insulation affected measurement accuracy. Sensors were first calibrated in a temperature bath (5-40°C) against a reference thermometer. Nineteen participants then completed two climate-chamber trials: a windy-chill condition (-7°C, 2.5 m/s wind) and cold-chill condition (-15°C, 0.2 m/s wind). Each sensor (iButtons®, BodyCAP, MSR, ADInstruments) was tested on the forearm with and without insulation. Infrared thermography (IRT) served as reference. In the water-bath comparison, all sensors performed within acceptable accuracy ranges, with biases from -0.13 to 0.29°C and 95% limits of agreement ≤ 0.63°C. During human trials, all sensors showed systemic overestimation, most pronounced for BodyCAP. Uninsulated, iButtons® overestimated temperatures relative to IRT in cold-chill and underestimated them in windy-chill conditions, while BodyCAP, MSR, and ADInstruments were unaffected by condition. Insulation consistently led to an overestimation of temperatures relative to IRT (bare bias: -1.24-2.84 °C, insulated bias: 1.41-7.5°C, P≤0.004). All sensors demonstrated temperature-dependent bias in the windy-chill condition, with larger deviations at higher skin temperatures. Error distributions were not significantly different over insulation or wind condition. Although all sensors showed acceptable accuracy in controlled water-bath calibration, systematic biases emerged during human trials. Insulation consistently caused overestimation, and iButtons® were sensitive to wind, whereas other sensors were not. Bias increased at higher skin temperatures under windy conditions across all sensor types.

Schlagwörter

Stichwörter der Autor:innen

coldcontact thermometryskin temperaturethermoregulationwindchill

Autor:innen und Institutionen

  • Human Performance, TNO, Kampweg 55, 3769DE Soesterberg, Soesterberg, 3769DE, Netherlands.
  • TNO, Kampweg 55, Soesterberg, 3769DE, Netherlands.
  • TNO, Kampweg 55, 3769DE Soesterberg, Soesterberg, ZH, 3769DE, Netherlands.

ORCID-Profile: Lisa Klous, Pim Scheiberlich, Aukje de Vrijer, Minke Pien Jolien Geerts, Anne Lamers

Förderung

  • Dutch Ministry of Defense

Zitieren

APA

Klous, L., Scheiberlich, P., de Vrijer, A., Geerts, M. P. J., Lamers, A., Van Der Sanden, K., & Kingma, B. (2026). Accuracy of Skin Temperature Sensors Under Cold and Windy Conditions: Effects of Sensor Type and Insulation. Physiological Measurement. https://doi.org/10.1088/1361-6579/aec1b2

Vancouver

Klous L, Scheiberlich P, de Vrijer A, Geerts MPJ, Lamers A, Van Der Sanden K, et al. Accuracy of Skin Temperature Sensors Under Cold and Windy Conditions: Effects of Sensor Type and Insulation. Physiological Measurement. 2026. doi:10.1088/1361-6579/aec1b2

BibTeX
@article{Klous2026,
  author = {Klous, Lisa and Scheiberlich, Pim and de Vrijer, Aukje and Geerts, Minke Pien Jolien and Lamers, Anne and Van Der Sanden, Koen and Kingma, Boris},
  title = {Accuracy of Skin Temperature Sensors Under Cold and Windy Conditions: Effects of Sensor Type and Insulation},
  journal = {Physiological Measurement},
  year = {2026},
  doi = {10.1088/1361-6579/aec1b2}
}
Kostenlos starten →