The MSPTDfast photoplethysmography beat detection algorithm: design, benchmarking, and open-source distribution
Physiological Measurement · 2025
DOI: 10.1088/1361-6579/adb89e ↗Kurz gefasst
Fachartikel in der Zeitschrift Physiological Measurement, 2025. Verfasst von Peter H Charlton und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Public Health and Primary Care, University of Cambridge, Cambridge, United Kingdom. Hauptthemen laut MEDLINE: Photoplethysmography, Algorithms, Signal Processing, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: British Heart Foundation.
Abstract (Originalsprache)
Objective: photoplethysmography is widely used for physiological monitoring, whether in clinical devices such as pulse oximeters, or consumer devices such as smartwatches. A key step in the analysis of photoplethysmogram (PPG) signals is detecting heartbeats. The multi-scale peak & trough detection ( MSPTD ) algorithm has been found to be one of the most accurate PPG beat detection algorithms, but is less computationally efficient than other algorithms. Therefore, the aim of this study was to develop a more efficient, open-source implementation of the MSPTD algorithm for PPG beat detection, named MSPTDfast (v.2) . Approach. five potential improvements to MSPTD were identified and evaluated on four datasets. MSPTDfast (v.2) was designed by incorporating each improvement which on its own reduced execution time whilst maintaining a high F 1 -score. After internal validation, MSPTDfast (v.2) was benchmarked against state-of-the-art beat detection algorithms on four additional datasets. Main results. MSPTDfast (v.2) incorporated two key improvements: pre-processing PPG signals to reduce the sampling frequency to 20 Hz; and only calculating scalogram scales corresponding to heart rates >30 bpm. During internal validation MSPTDfast (v.2) was found to have an execution time of between approximately one-third and one-twentieth of MSPTD , and a comparable F 1 -score. During benchmarking MSPTDfast (v.2) was found to have the highest F 1 -score alongside MSPTD , and amongst one of the lowest execution times with only MSPTDfast (v.1) , qppgfast and MMPD (v.2) achieving shorter execution times. Significance. MSPTDfast (v.2) is an accurate and efficient PPG beat detection algorithm, available in an open-source Matlab toolbox.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Public Health and Primary Care, University of Cambridge, Cambridge, United Kingdom.
- Research Centre of Biomedical Engineering, City, University of London, London, United Kingdom.
- Programa de Bioingeniería, Facultad de Ingeniería, Universidad Santiago de Cali, Cali, Colombia.
ORCID-Profile: Peter H Charlton, Erick Javier Argüello-Prada, Jonathan Mant, Panicos A Kyriacou
Förderung
- British Heart Foundation
Zitieren
APA
Charlton, P. H., Argüello-Prada, E. J., Mant, J., & Kyriacou, P. A. (2025). The MSPTDfast photoplethysmography beat detection algorithm: design, benchmarking, and open-source distribution. Physiological Measurement. https://doi.org/10.1088/1361-6579/adb89e
Vancouver
Charlton PH, Argüello-Prada EJ, Mant J, Kyriacou PA. The MSPTDfast photoplethysmography beat detection algorithm: design, benchmarking, and open-source distribution. Physiological Measurement. 2025. doi:10.1088/1361-6579/adb89e
BibTeX
@article{Charlton2025,
author = {Charlton, Peter H and Argüello-Prada, Erick Javier and Mant, Jonathan and Kyriacou, Panicos A},
title = {The MSPTDfast photoplethysmography beat detection algorithm: design, benchmarking, and open-source distribution},
journal = {Physiological Measurement},
year = {2025},
doi = {10.1088/1361-6579/adb89e}
}
EUVIMED