Multimodal wearable EEG, EMG and accelerometry measurements improve the accuracy of tonic-clonic seizure detection
Physiological Measurement · 01.06.2024 · 45 (6) · 065005
DOI: 10.1088/1361-6579/ad4e94 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physiological Measurement, online erschienen am 07.06.2024. Verfasst von Jingwei Zhang und 25 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering (ESAT), STADIUS Center for Dynamical Systems, Signal Processing and Data Analytics, KU Leuven, Leuven, Belgium. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Electroencephalography, Electromyography, Accelerometry. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 51 Literaturangaben. Gefördert von: EIT Health, FWO Research Project, Flemish Government.
Abstract (Originalsprache)
Objective
This paper aims to investigate the possibility of detecting tonic-clonic seizures (TCSs) with behind-the-ear, two-channel wearable electroencephalography (EEG), and to evaluate its added value to non-EEG modalities in TCS detection.
Methods
We included 27 participants with a total of 44 TCSs from the European multicenter study SeizeIT2. The wearable Sensor Dot (Byteflies) was used to measure behind-the-ear EEG, electromyography (EMG), electrocardiography, accelerometry (ACC) and gyroscope. We evaluated automatic unimodal detection of TCSs, using sensitivity, precision, false positive rate (FPR) and F1-score. Subsequently, we fused the different modalities and again assessed performance. Algorithm-labeled segments were then provided to two experts, who annotated true positive TCSs, and discarded false positives.
Results
Wearable EEG outperformed the other single modalities with a sensitivity of 100% and a FPR of 10.3/24 h. The combination of wearable EEG and EMG proved most clinically useful, delivering a sensitivity of 97.7%, an FPR of 0.4/24 h, a precision of 43%, and an F1-score of 59.7%. The highest overall performance was achieved through the fusion of wearable EEG, EMG, and ACC, yielding a sensitivity of 90.9%, an FPR of 0.1/24 h, a precision of 75.5%, and an F1-score of 82.5%.
Conclusions
In TCS detection with a wearable device, combining EEG with EMG, ACC or both resulted in a remarkable reduction of FPR, while retaining a high sensitivity. Significance . Adding wearable EEG could further improve TCS detection, relative to extracerebral-based systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical Engineering (ESAT), STADIUS Center for Dynamical Systems, Signal Processing and Data Analytics, KU Leuven, Leuven, Belgium.
- Laboratory for Epilepsy Research, KU Leuven, Leuven, Belgium.
- Reference Center for Refractory epilepsy, Department of Neurology, UZ Leuven, Leuven, Belgium.
- Department of Pediatric Neurology, UZ Leuven, Leuven, Belgium.
- Department of Development and Regeneration, KU Leuven, Leuven, Belgium.
- Department of Neurology, Karolinska University Hospital, Stockholm, Sweden.
ORCID-Profile: Jingwei Zhang, Lauren Swinnen, Christos Chatzichristos, Nicolas Zabler, Matthias Dümpelmann
Förderung
- EIT Health
- FWO Research Project
- Flemish Government
- KU Leuven
Zitieren
APA
Zhang, J., Swinnen, L., Chatzichristos, C., Broux, V., Proost, R., Jansen, K., Mahler, B., Zabler, N., Epitashvilli, N., Dümpelmann, M., Schulze-Bonhage, A., Schriewer, E., Ermis, U., Wolking, S., Linke, F., Weber, Y., Symmonds, M., Sen, A., Biondi, A., …, & Vos, M. (2024). Multimodal wearable EEG, EMG and accelerometry measurements improve the accuracy of tonic-clonic seizure detection. Physiological Measurement, 45(6), 065005. https://doi.org/10.1088/1361-6579/ad4e94
Vancouver
Zhang J, Swinnen L, Chatzichristos C, Broux V, Proost R, Jansen K, et al. Multimodal wearable EEG, EMG and accelerometry measurements improve the accuracy of tonic-clonic seizure detection. Physiological Measurement. 2024;45(6):065005. doi:10.1088/1361-6579/ad4e94
BibTeX
@article{Zhang2024,
author = {Zhang, Jingwei and Swinnen, Lauren and Chatzichristos, Christos and Broux, Victoria and Proost, Renee and Jansen, Katrien and Mahler, Benno and Zabler, Nicolas and Epitashvilli, Nino and Dümpelmann, Matthias and Schulze-Bonhage, Andreas and Schriewer, Elisabeth and Ermis, Ummahan and Wolking, Stefan and Linke, Florian and Weber, Yvonne and Symmonds, Mkael and Sen, Arjune and Biondi, Andrea and Richardson, Mark P and I, Abuhaiba Sulaiman and Silva, Ana Isabel and Sales, Francisco and Vértes, Gergely and Paesschen, Wim Van and Vos, Maarten De},
title = {Multimodal wearable EEG, EMG and accelerometry measurements improve the accuracy of tonic-clonic seizure detection},
journal = {Physiological Measurement},
year = {2024},
volume = {45},
number = {6},
pages = {065005},
doi = {10.1088/1361-6579/ad4e94}
}
EUVIMED