EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Scatter-induced non-linear signal compression and SDNR loss in triple-energy photon-counting iodine imaging

Jesse Tanguay iD, Kaitlyn Sims iD

Physics in Medicine & Biology · 2026

DOI: 10.1088/1361-6560/aeafa4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2026. Verfasst von Jesse Tanguay und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Toronto Metropolitan University, Toronto, Ontario, Canada. Hauptthemen laut MEDLINE: Photons, Iodine, Scattering, Radiation, Signal-To-Noise Ratio. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Canada Foundation for Innovation, Natural Sciences and Engineering Research Council of Canada.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective. Triple-energy x-ray imaging with photon-counting detectors (PCDs) enables material decomposition, but energy-dependent scatter introduces bin-specific distortions that propagate nonlinearly through triple-energy subtraction. This work quantifies how bin-dependent scatter alters iodine signal formation and affects signal-difference-to-noise ratio (SDNR). Approach. Analytical expressions for iodine signal difference, noise and SDNR were derived incorporating energy-bin-specific scatter fractions (SF) and their nonlinear coupling with iodine areal density. Measurements used a CdTe-based PCD and an image-quality phantom, with scatter varied through field of view and quantified using the air-gap method. SDNR was evaluated over iodine areal densities of 33–140 mg cm −2 , SF values of 0.2–0.8, tube voltages of 60–100 kV and phantom thicknesses of 10–20 cm. A physics-informed empirical model was fitted to the data. Simulations incorporating charge sharing and beam hardening supported interpretation. Main results. Scatter fractions differed across energy bins, with the lowest-energy bin exhibiting the highest SF. Increasing scatter introduced curvature in energy-bin iodine signals versus areal density. Through triple-energy subtraction, these unequal responses produced pronounced signal compression. The model captured energy-bin and combined signal differences (R 2 > 0.92), with reduced agreement for SDNR (R 2 = 0.82–0.94), reflecting additional coupling between signal, noise and scatter. SDNR degradation exceeded first-order scatter predictions: for 52 mg cm −2 iodine, SDNR decreased by 20% at SF ≈ 0.2 and 70% at SF ≈ 0.8. Simulations reproduced both linear contrast loss and additional nonlinear compression. Significance. Bin-dependent scatter induces nonlinear signal compression that amplifies SDNR loss with increasing iodine areal density. Because this behavior arises from the structure of triple-energy subtraction, first-order scatter scaling systematically underestimates performance degradation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhotonsIodineScattering, RadiationSignal-To-Noise RatioNonlinear DynamicsImage Processing, Computer-AssistedPhantoms, Imaging

Stichwörter der Autor:innen

image qualityphoton-counting detector imagingsignal-to-noise ratiospectral x-ray imagingtriple-energy subtractionvascular imagingx-ray scatter

Substanzen

Iodine

Autor:innen und Institutionen

  • Department of Physics, Toronto Metropolitan University, Toronto, Ontario, Canada.

ORCID-Profile: Jesse Tanguay, Kaitlyn Sims

Förderung

  • Canada Foundation for Innovation
  • Natural Sciences and Engineering Research Council of Canada

Zitieren

APA

Tanguay, J., & Sims, K. (2026). Scatter-induced non-linear signal compression and SDNR loss in triple-energy photon-counting iodine imaging. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/aeafa4

Vancouver

Tanguay J, Sims K. Scatter-induced non-linear signal compression and SDNR loss in triple-energy photon-counting iodine imaging. Physics in Medicine & Biology. 2026. doi:10.1088/1361-6560/aeafa4

BibTeX
@article{Tanguay2026,
  author = {Tanguay, Jesse and Sims, Kaitlyn},
  title = {Scatter-induced non-linear signal compression and SDNR loss in triple-energy photon-counting iodine imaging},
  journal = {Physics in Medicine & Biology},
  year = {2026},
  doi = {10.1088/1361-6560/aeafa4}
}
Kostenlos starten →