Analytical model of large-diameter spherical-like catheter for pulsed field ablation in treating cardiac arrhythmias
Physics in Medicine & Biology · 2025
DOI: 10.1088/1361-6560/ae1bf3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2025. Verfasst von Juan J Pérez und 2 weiteren Autor:innen. Beteiligte Einrichtung: BioMIT, Department of Electronic Engineering, Universitat Politècnica de València, Valencia, Spain. Hauptthemen laut MEDLINE: Arrhythmias, Cardiac, Catheter Ablation, Catheters. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Spanish Ministerio de Ciencia e Innovación, Agencia Estatal de Investigación, Fondo Europeo de Desarrollo Regional.
Abstract (Originalsprache)
Objective. To study the electrical performance of large-diameter spherical-like catheters used for pulsed field ablation (PFA) in treating cardiac arrhythmias, especially evaluating the effect of electrode radius, applied voltage and maximum current delivered on the PFA-induced lesion depth. Approach. An analytical model was proposed which used spherical coordinates to estimate the value of the electric field around a spherical metal electrode. Lesion depth was estimated by the myocardium lethal electric field threshold. The paper also reviews the literature and proposes a relationship between this threshold and the total cumulative duration of the pulses applied during PFA. Main results. Despite the model´s simplicity in geometric terms, the estimated values of baseline impedance (44 Ω) and impedance drop due to PFA (7–9 Ω) are closely related to those observed in experimental studies (43 Ω and 5–7 Ω, respectively). The PFA-induced lesion size potentially increases with the applied voltage, but this is only true if the generator can provide the electrical current needed to maintain the required voltage. A nonlinear relationship was identified between PFA-induced lesion depth and the electrode diameter, which was also modulated by the total duration. Significance. The analytical model offers a physical explanation for the PFA-Induced electrical phenomenon, while emphasising how certain technical characteristics (maximum current, applied voltage, electrode diameter) can affect the depth of the lesion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- BioMIT, Department of Electronic Engineering, Universitat Politècnica de València, Valencia, Spain.
- Electrophysiology Unit of Cardiology Department, Hospital Universitario y Politécnico La Fe de Valencia, Valencia, Spain.
ORCID-Profile: Juan J Pérez, Ana González-Suárez
Förderung
- Spanish Ministerio de Ciencia e Innovación
- Agencia Estatal de Investigación
- Fondo Europeo de Desarrollo Regional
- Spanish Ministerio de Ciencia e Innovación, Agencia Estatal de Investigación, Fondo Europeo de Desarrollo Regional
Zitieren
APA
Pérez, J. J., Izquierdo, M., & González-Suárez, A. (2025). Analytical model of large-diameter spherical-like catheter for pulsed field ablation in treating cardiac arrhythmias. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ae1bf3
Vancouver
Pérez JJ, Izquierdo M, González-Suárez A. Analytical model of large-diameter spherical-like catheter for pulsed field ablation in treating cardiac arrhythmias. Physics in Medicine & Biology. 2025. doi:10.1088/1361-6560/ae1bf3
BibTeX
@article{Prez2025,
author = {Pérez, Juan J and Izquierdo, Maite and González-Suárez, Ana},
title = {Analytical model of large-diameter spherical-like catheter for pulsed field ablation in treating cardiac arrhythmias},
journal = {Physics in Medicine & Biology},
year = {2025},
doi = {10.1088/1361-6560/ae1bf3}
}