Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators
Physics in Medicine & Biology · 2025
DOI: 10.1088/1361-6560/addfa7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2025. Verfasst von Ioannis Androulakis und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center, Rotterdam, The Netherlands. Hauptthemen laut MEDLINE: Brachytherapy, Electromagnetic Phenomena, Genital Neoplasms, Female, Quality Assurance, Health Care. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Elekta AB, Stockholm, Sweden.
Abstract (Originalsprache)
Objective. In brachytherapy for gynecological cancers using intracavitary applicators, implant reconstruction is commonly performed using applicator libraries. These libraries contain applicator geometry models as well as dwell position (DP) models defined in respect to the applicator geometry. In this study, we investigate whether an afterloader integrated electromagnetic tracking (EMT) system can be utilized for DP model definition and quality assurance in such applicators. Approach. DPs in four sets of two configurations of the Elekta Venezia Advanced Gynaecological Applicator (22 mm ovoids/40 mm intrauterine (IU) and 26 mm ovoids/70 mm IU) were measured using an afterloader integrated EMT system. Measurements were evaluated for reproducibility and compared against manufacturer-specified (MS) DPs and a computed tomography (CT)-corrected DP model. Main Results. Excellent EMT measurement reproducibility was observed, with values of ⩽0.2 mm for both configurations. The overall reproducibility, including applicator geometry reproducibility, was ⩽0.4 mm for both configurations. Significant discrepancies from the manufacturer’s DP model were observed, with a mean ± sd deviation of 1.13 ± 0.66 mm (22/40) and 1.37 ± 0.63 (26/70), particularly in the IU channel, where MS DPs were not experimentally defined. Discrepancies were reduced to 0.89 ± 0.41 mm (22/40) and 0.81 ± 0.33 mm (26/70) when the CT-corrected DP model was used as baseline, highlighting the need for experimentally defined DP models. The overall uncertainty of single measurements was below the clinically acceptable 2 mm limit. Significance. This study confirms that afterloader integrated EMT can accurately reconstruct source paths in gynecological brachytherapy applicators and supports its incorporation into clinical workflows for improved quality assurance and treatment precision. The importance of EMT for quality assurance was highlighted by measured deviations from manufacturer’s DP model in a clinical relevant part of the IU channel.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center, Rotterdam, The Netherlands.
- Department of Medical Physics and Informatics, HollandPTC, Delft, The Netherlands.
- Department of BioMechanical Engineering, Delft University of Technology, Delft, The Netherlands.
ORCID-Profile: Ioannis Androulakis, Jeremy Godart, Robin Straathof, Henrike Westerveld, Remi Nout, Mischa Hoogeman, Inger-Karine K Kolkman-Deurloo
Förderung
- Elekta AB, Stockholm, Sweden
Zitieren
APA
Androulakis, I., van Laar, M., Godart, J., Straathof, R., Westerveld, H., Nout, R., Hoogeman, M., & Kolkman-Deurloo, I. K. (2025). Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/addfa7
Vancouver
Androulakis I, van Laar M, Godart J, Straathof R, Westerveld H, Nout R, et al. Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators. Physics in Medicine & Biology. 2025. doi:10.1088/1361-6560/addfa7
BibTeX
@article{Androulakis2025,
author = {Androulakis, Ioannis and van Laar, Myra and Godart, Jeremy and Straathof, Robin and Westerveld, Henrike and Nout, Remi and Hoogeman, Mischa and Kolkman-Deurloo, Inger-Karine K},
title = {Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators},
journal = {Physics in Medicine & Biology},
year = {2025},
doi = {10.1088/1361-6560/addfa7}
}
EUVIMED