A potential usefulness of ultra-high-resolution computed tomography in quality assurance of remote after-loading system for cervical cancer
Journal of Radiation Research · 2024
DOI: 10.1093/jrr/rrae055 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Radiation Research, 2024. Verfasst von Masashi Kinjyo und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Graduate School of Medical Science, University of the Ryukyus, 207 Uehara, Nishihara-cho, Nakagami-gun, Okinawa 903-0215, Ja…. Hauptthemen laut MEDLINE: Uterine Cervical Neoplasms, Tomography, X-Ray Computed, Quality Assurance, Health Care, Brachytherapy. Die Arbeit nennt 7 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Canon Medical Systems.
Abstract (Originalsprache)
Intracavitary brachytherapy with a remote after-loading system (RALS) is performed as a part of radical radiation therapy in cervical cancer. The radiation source is delivered directly through an applicator placed inside the uterus or vagina. Thorough quality control is important to prevent accidents that can lead to serious irradiation error, and an applicator check is one such quality control measure. We experienced a clinical situation in which a small volume of water was observed in the lumen of a post-sterilized applicator on treatment-planning CT. Although the submersion test was negative and no air bubbles emerged from the applicator, ultra-high-resolution computed tomography (U-HRCT) showed a linear crack reaching the inside of the applicator. This abnormality was not identified on treatment-planning CT, which has lower spatial resolution than U-HRCT. In addition, no linear cracks were seen on U-HRCT images of eight other applicators considered to be free from damage. U-HRCT may have superior potential to detect applicator damage and could be useful for quality assurance of the RALS procedure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Graduate School of Medical Science, University of the Ryukyus, 207 Uehara, Nishihara-cho, Nakagami-gun, Okinawa 903-0215, Japan.
- Division of Radiology, University of the Ryukyus Hospital, 207 Uehara, Nishihara-cho, Nakagami-gun, Okinawa 903-0215, Japan.
- Canon Medical Systems Corporation, 70-1 Yanagi-cho, Saiwai-ku, Kawasaki, Kanagawa 212-0015, Japan.
Förderung
- Canon Medical Systems
Zitieren
APA
Kinjyo, M., Nishie, A., Kudaka, R., Nakano, S., & Ariga, T. (2024). A potential usefulness of ultra-high-resolution computed tomography in quality assurance of remote after-loading system for cervical cancer. Journal of Radiation Research. https://doi.org/10.1093/jrr/rrae055
Vancouver
Kinjyo M, Nishie A, Kudaka R, Nakano S, Ariga T. A potential usefulness of ultra-high-resolution computed tomography in quality assurance of remote after-loading system for cervical cancer. Journal of Radiation Research. 2024. doi:10.1093/jrr/rrae055
BibTeX
@article{Kinjyo2024,
author = {Kinjyo, Masashi and Nishie, Akihiro and Kudaka, Ryo and Nakano, Shota and Ariga, Takuro},
title = {A potential usefulness of ultra-high-resolution computed tomography in quality assurance of remote after-loading system for cervical cancer},
journal = {Journal of Radiation Research},
year = {2024},
doi = {10.1093/jrr/rrae055}
}
EUVIMED