Virtual dosimetry study with three cone-beam breast computed tomography scanners using a fast GPU-based Monte Carlo code
Physics in Medicine & Biology · 2024
DOI: 10.1088/1361-6560/ad2012 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2024. Verfasst von Giovanni Mettivier und 3 weiteren Autor:innen. Beteiligte Einrichtung: Dipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, I-80126 Naples, Italy. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Radiometry. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 77 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Istituto Nazionale di Fisica Nucleare (INFN).
Abstract (Originalsprache)
Objective . To compare the dosimetric performance of three cone-beam breast computed tomography (BCT) scanners, using real-time Monte Carlo-based dose estimates obtained with the virtual clinical trials (VCT)-BREAST graphical processing unit (GPU)-accelerated platform dedicated to VCT in breast imaging . Approach . A GPU-based Monte Carlo (MC) code was developed for replicating in silico the geometric, x-ray spectra and detector setups adopted, respectively, in two research scanners and one commercial BCT scanner, adopting 80 kV, 60 kV and 49 kV tube voltage, respectively. Our cohort of virtual breasts included 16 anthropomorphic voxelized breast phantoms from a publicly available dataset. For each virtual patient, we simulated exams on the three scanners, up to a nominal simulated mean glandular dose of 5 mGy (primary photons launched, in the order of 10 11 –10 12 per scan). Simulated 3D dose maps (recorded for skin, adipose and glandular tissues) were compared for the same phantom, on the three scanners. MC simulations were implemented on a single NVIDIA GeForce RTX 3090 graphics card. Main results. Using the spread of the dose distribution as a figure of merit, we showed that, in the investigated phantoms, the glandular dose is more uniform within less dense breasts, and it is more uniformly distributed for scans at 80 kV and 60 kV, than at 49 kV. A realistic virtual study of each breast phantom was completed in about 3.0 h with less than 1% statistical uncertainty, with 10 9 primary photons processed in 3.6 s computing time. Significance . We reported the first dosimetric study of the VCT-BREAST platform, a fast MC simulation tool for real-time virtual dosimetry and imaging trials in BCT, investigating the dose delivery performance of three clinical BCT scanners. This tool can be adopted to investigate also the effects on the 3D dose distribution produced by changes in the geometrical and spectrum characteristics of a cone-beam BCT scanner.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Dipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, I-80126 Naples, Italy.
- INFN Sezione di Napoli, I-80126 Naples, Italy.
- Innovative Technology of Radiotherapy Computation and Hardware (iTORCH) Laboratory, Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX 752878, United States of America.
- Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD 21224, United States of America.
ORCID-Profile: Giovanni Mettivier, Youfang Lai, Xun Jia, Paolo Russo
Förderung
- Istituto Nazionale di Fisica Nucleare (INFN)
Zitieren
APA
Mettivier, G., Lai, Y., Jia, X., & Russo, P. (2024). Virtual dosimetry study with three cone-beam breast computed tomography scanners using a fast GPU-based Monte Carlo code. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ad2012
Vancouver
Mettivier G, Lai Y, Jia X, Russo P. Virtual dosimetry study with three cone-beam breast computed tomography scanners using a fast GPU-based Monte Carlo code. Physics in Medicine & Biology. 2024. doi:10.1088/1361-6560/ad2012
BibTeX
@article{Mettivier2024,
author = {Mettivier, Giovanni and Lai, Youfang and Jia, Xun and Russo, Paolo},
title = {Virtual dosimetry study with three cone-beam breast computed tomography scanners using a fast GPU-based Monte Carlo code},
journal = {Physics in Medicine & Biology},
year = {2024},
doi = {10.1088/1361-6560/ad2012}
}
EUVIMED