Sub-nanosecond measurements of high-current and high-voltage pulses
Review of Scientific Instruments · 01.08.2026 · 97 (8)
DOI: 10.1063/5.0335830 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 14.08.2026. Verfasst von Yan Zeltser und 3 weiteren Autor:innen. Beteiligte Einrichtung: Electrical Engineering Department, Technion - Israel Institute of Technology, Haifa 32000, Israel. Die Arbeit nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
We present a measurement tool encompassing four different types of sensors: a B-dot, a D-dot, a capacitive voltage divider, and a low-inductance current-view resistor. The tool can measure high-voltage, high-current pulses with rise times <150 ps. A calibration procedure for the device using a commercial pulse generator is demonstrated. Measurements of pulses with <150 ps rise time generated by a spiral generator with sharpening spark gaps are also presented.
Autor:innen und Institutionen
- Electrical Engineering Department, Technion - Israel Institute of Technology, Haifa 32000, Israel.
- Physics Department, Technion - Israel Institute of Technology, Haifa 32000, Israel.
ORCID-Profile: Yan Zeltser, Viktor Akhmetzianov, Levi Schächter, Yakov E Krasik
Zitieren
APA
Zeltser, Y., Akhmetzianov, V., Schächter, L., & Krasik, Y. E. (2026). Sub-nanosecond measurements of high-current and high-voltage pulses. Review of Scientific Instruments, 97(8). https://doi.org/10.1063/5.0335830
Vancouver
Zeltser Y, Akhmetzianov V, Schächter L, Krasik YE. Sub-nanosecond measurements of high-current and high-voltage pulses. Review of Scientific Instruments. 2026;97(8). doi:10.1063/5.0335830
BibTeX
@article{Zeltser2026,
author = {Zeltser, Yan and Akhmetzianov, Viktor and Schächter, Levi and Krasik, Yakov E},
title = {Sub-nanosecond measurements of high-current and high-voltage pulses},
journal = {Review of Scientific Instruments},
year = {2026},
volume = {97},
number = {8},
doi = {10.1063/5.0335830}
}