EUVIMEDEUVIMED entdecken
Crossref

A compact low-power magnetic particle imaging scanner based on a permanent-magnet field-free-line generator with high gradient

Zechen Wei iD, Tao Zhu iD, Bingye Wang iD, Runze Li iD, Jie Tian iD, Hui Hui iD

Review of Scientific Instruments · 01.06.2026 · 97 (6)

DOI: 10.1063/5.0324735 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 25.06.2026. Verfasst von Zechen Wei und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Automation, Chinese Academy of Sciences, Beijing 100190, China. Die Arbeit nennt 40 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Young Talent Support Project of Guangzhou Association for Science and Technology, Beijing Natural Science Foundation.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Field-free-line (FFL) configurations have the potential to enhance the sensitivity of magnetic particle imaging (MPI). However, electromagnet-based implementations typically suffer from high power consumption, limiting scalability toward human-sized systems, while permanent-magnet-based approaches often require large volume and weight to generate high-gradient and uniform FFLs. In this work, we present a compact permanent-magnet assembly that generates a uniform, high-gradient FFL without power consumption. The system employs four magnet pairs to produce a selection field gradient of up to 4.068 T/m/μ0 while maintaining a reduced footprint. To reduce system complexity and enrich spectral content, a single-coil, coaxial multi-frequency excitation scheme is adopted. In addition, a multi-angle FFL rotation imaging strategy combined with a system-matrix-based reconstruction framework is implemented to enhance spatial encoding and reconstruction fidelity. Magnetic field calibration confirms the generation of a stable and uniform FFL using magnet pairs with dimensions of 48 × 100 × 150 mm3. Comparative simulations demonstrate that, relative to conventional permanent-magnet designs, the proposed configuration reduces magnet volume by one order of magnitude, while completely eliminating power consumption compared to electromagnet-based systems requiring kilowatt-level power. Imaging experiments further demonstrate a spatial resolution of 0.3 mm along the main gradient direction and an imaging sensitivity of 98 ng, with reconstruction quality improving as angular coverage increases. Finally, two fully acquired multi-angle FFL system matrices and corresponding phantom datasets are open-sourced. Overall, this work provides a practical hardware solution for high-performance FFL-MPI and offers benchmark datasets to support algorithm development and system optimization toward clinically relevant MPI applications.

Autor:innen und Institutionen

  • Institute of Automation, Chinese Academy of Sciences, Beijing 100190, China.
  • Beijing Key Laboratory of Molecular Imaging Technology Research and Translation, Beijing 100190, China.
  • Department of Biomedical Engineering, The Hong Kong Polytechnic University, Hong Kong 999077, China.
  • Research Center for Medical AI, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen 518067, China.
  • School of Artificial Intelligence, University of Chinese Academy of Sciences, Beijing 100080, China.
  • School of Engineering Medicine and School of Biological Science and Medical Engineering, Beihang University, Beijing 100191, China.

ORCID-Profile: Zechen Wei, Tao Zhu, Bingye Wang, Runze Li, Jie Tian, Hui Hui

Förderung

  • National Natural Science Foundation of China
  • Young Talent Support Project of Guangzhou Association for Science and Technology
  • Beijing Natural Science Foundation
  • CAS Youth Innovation Promotion Association
  • Noncommunicable Chronic Diseases-National Science and Technology Major Project of China

Zitieren

APA

Wei, Z., Zhu, T., Wang, B., Li, R., Tian, J., & Hui, H. (2026). A compact low-power magnetic particle imaging scanner based on a permanent-magnet field-free-line generator with high gradient. Review of Scientific Instruments, 97(6). https://doi.org/10.1063/5.0324735

Vancouver

Wei Z, Zhu T, Wang B, Li R, Tian J, Hui H. A compact low-power magnetic particle imaging scanner based on a permanent-magnet field-free-line generator with high gradient. Review of Scientific Instruments. 2026;97(6). doi:10.1063/5.0324735

BibTeX
@article{Wei2026,
  author = {Wei, Zechen and Zhu, Tao and Wang, Bingye and Li, Runze and Tian, Jie and Hui, Hui},
  title = {A compact low-power magnetic particle imaging scanner based on a permanent-magnet field-free-line generator with high gradient},
  journal = {Review of Scientific Instruments},
  year = {2026},
  volume = {97},
  number = {6},
  doi = {10.1063/5.0324735}
}