Multi-delay coherence imaging spectroscopy optimized for ion temperature measurements in the divertor plasma of the Wendelstein 7-X stellarator
Review of Scientific Instruments · 01.07.2024 · 95 (7)
DOI: 10.1063/5.0208586 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 01.07.2024. Verfasst von David M Kriete und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Auburn University, Auburn, Alabama 36849, USA. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: U.S. Department of Energy, EUROfusion.
Abstract (Originalsprache)
A new coherence imaging spectroscopy (CIS) diagnostic optimized to measure the C2+ impurity ion temperature Ti spatial distribution in the divertor plasma of the W7-X stellarator is designed, tested, and validated. Using CIS to obtain Ti in the edge of magnetically confined plasmas has historically been challenging because Doppler broadening and Zeeman splitting have comparable effects on the shape of spectral emission lines. To distinguish between these two mechanisms, a novel approach to birefringent crystal design is employed to minimize the diagnostic’s sensitivity to Zeeman splitting. The recently developed pixelated multi-delay CIS approach is also used to obtain four times as much spectral information as traditional CIS approaches. The Ti-optimized CIS diagnostic is validated in a long-pulse W7-X plasma by comparison with a high-resolution spectrometer whose sightlines overlap with the CIS field of view. The CIS and spectrometer Ti profiles have the same shape and agree to within 10% on average and 25% in the worst case. Images of the Ti distribution near the divertor show toroidally elongated bands aligned with the magnetic field, with Ti ranging between 10 and 40 eV.
Autor:innen und Institutionen
- Department of Physics, Auburn University, Auburn, Alabama 36849, USA.
- Max-Planck-Institut für Plasmaphysik, Greifswald, Mecklenburg-Vorpommern 17491, Germany.
ORCID-Profile: David M Kriete, Valeria Perseo, Dorothea Gradic, David A Ennis, Ralf König, David A Maurer
Förderung
- U.S. Department of Energy
- EUROfusion
Zitieren
APA
Kriete, D. M., Perseo, V., Gradic, D., Ennis, D. A., König, R., Maurer, D. A., & W7-X Team (2024). Multi-delay coherence imaging spectroscopy optimized for ion temperature measurements in the divertor plasma of the Wendelstein 7-X stellarator. Review of Scientific Instruments, 95(7). https://doi.org/10.1063/5.0208586
Vancouver
Kriete DM, Perseo V, Gradic D, Ennis DA, König R, Maurer DA, et al. Multi-delay coherence imaging spectroscopy optimized for ion temperature measurements in the divertor plasma of the Wendelstein 7-X stellarator. Review of Scientific Instruments. 2024;95(7). doi:10.1063/5.0208586
BibTeX
@article{Kriete2024,
author = {Kriete, David M and Perseo, Valeria and Gradic, Dorothea and Ennis, David A and König, Ralf and Maurer, David A and W7-X Team, W7-X Team},
title = {Multi-delay coherence imaging spectroscopy optimized for ion temperature measurements in the divertor plasma of the Wendelstein 7-X stellarator},
journal = {Review of Scientific Instruments},
year = {2024},
volume = {95},
number = {7},
doi = {10.1063/5.0208586}
}
EUVIMED