New dynamic diamond anvil cell for time-resolved radial x-ray diffraction
Review of Scientific Instruments · 01.04.2024 · 95 (4)
DOI: 10.1063/5.0179799 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 17.04.2024. Verfasst von L Q Huston und 7 weiteren Autor:innen. Beteiligte Einrichtung: Dynamic Experiments Division, Los Alamos National Laboratory, Los Alamos, New Mexico 87545, USA. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 67 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science Foundation, U.S. Department of Energy.
Abstract (Originalsprache)
The dynamic diamond anvil cell (dDAC) is a recently developed experimental platform that has shown promise for studying the behavior of materials at strain rates ranging from intermediate to quasi-static and shock compression regimes. Combining dDAC with time-resolved x-ray diffraction (XRD) in the radial geometry (i.e., with incident x-rays perpendicular to the axis of compression) enables the study of material properties such as strength, texture evolution, and deformation mechanisms. This work describes a radial XRD dDAC setup at beamline P02.2 (Extreme Conditions Beamline) at DESY’s PETRA III synchrotron. Time-resolved radial XRD data are collected for titanium, zirconium, and zircon samples, demonstrating the ability to study the strength and texture of materials at compression rates above 300 GPa/s. In addition, the simultaneous optical imaging of the DAC sample chamber is demonstrated. The ability to conduct simultaneous radial XRD and optical imaging provides the opportunity to characterize plastic strain and deviatoric strain rates in the DAC at intermediate rates, exploring the strength and deformation mechanisms of materials in this regime.
Autor:innen und Institutionen
- Dynamic Experiments Division, Los Alamos National Laboratory, Los Alamos, New Mexico 87545, USA.
- Department of Geology and Geophysics, University of Utah, Salt Lake City, Utah 84112, USA.
- Deutsches Elektronen Synchrotron, Notkestr. 85, 22607 Hamburg, Germany.
ORCID-Profile: L Q Huston, L Miyagi, R J Husband, K Glazyrin, C Kiessner, M Wendt, H P Liermann, B T Sturtevant
Förderung
- National Science Foundation
- U.S. Department of Energy
Zitieren
APA
Huston, L. Q., Miyagi, L., Husband, R. J., Glazyrin, K., Kiessner, C., Wendt, M., Liermann, H. P., & Sturtevant, B. T. (2024). New dynamic diamond anvil cell for time-resolved radial x-ray diffraction. Review of Scientific Instruments, 95(4). https://doi.org/10.1063/5.0179799
Vancouver
Huston LQ, Miyagi L, Husband RJ, Glazyrin K, Kiessner C, Wendt M, et al. New dynamic diamond anvil cell for time-resolved radial x-ray diffraction. Review of Scientific Instruments. 2024;95(4). doi:10.1063/5.0179799
BibTeX
@article{Huston2024,
author = {Huston, L Q and Miyagi, L and Husband, R J and Glazyrin, K and Kiessner, C and Wendt, M and Liermann, H P and Sturtevant, B T},
title = {New dynamic diamond anvil cell for time-resolved radial x-ray diffraction},
journal = {Review of Scientific Instruments},
year = {2024},
volume = {95},
number = {4},
doi = {10.1063/5.0179799}
}
EUVIMED