Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 17.09.2024. Verfasst von N E Palmer und 11 weiteren Autor:innen. Beteiligte Einrichtung: Lawrence Livermore National Laboratory, Livermore, California 94550, USA. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 12 Literaturangaben. Gefördert von: U.S. Department of Energy.
Abstract (Originalsprache)
As part of a program to measure phase transition timescales in materials under dynamic compression, we have designed new x-ray imaging diagnostics to record multiple x-ray diffraction measurements during a single laser-driven experiment. Our design places several ns-gated hybrid CMOS (hCMOS) sensors within a few cm of a laser-driven target. The sensors must be protected from an extremely harsh environment, including debris, electromagnetic pulses, and unconverted laser light. Another key challenge is reducing the x-ray background relative to the faint diffraction signal. Building on the success of our predecessor (Target Diffraction In Situ), we implemented a staged approach to platform development. First, we built a demonstration diagnostic (Gated Diffraction Development Diagnostic) with two hCMOS sensors to confirm we could adequately protect them from the harsh environment and also acquire acceptable diffraction data. This allowed the team to quickly assess the risks and address the most significant challenges. We also collected scientifically useful data during development. Leveraging what we learned, we recently developed a much more ambitious instrument (Flexible Imaging Diffraction Diagnostic for Laser Experiments) that can field up to eight hCMOS sensors in a flexible geometry and participate in back-to-back shots at the National Ignition Facility (NIF). The design also allows for future iterations, such as faster hCMOS sensors and an embedded x-ray streak camera. The enhanced capabilities of the new instrument required a much more complex design, and the unexpected issues encountered on the first few shots at NIF remind us that complexity has consequences. Our progress in addressing these challenges is described herein, as is our current focus on improving data quality by reducing x-ray background and quantifying the uncertainties of our diffraction measurements.
Autor:innen und Institutionen
- Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
ORCID-Profile: N E Palmer, L R Benedetti, C E Vennari, P R Nyholm, N Bhandarkar, S R Nagel, J H Eggert, D K Bradley, A J Mackinnon, Y Ping
Förderung
- U.S. Department of Energy
Zitieren
APA
Palmer, N. E., Benedetti, L. R., Vennari, C. E., Nyholm, P. R., Petre, R. B., Bhandarkar, N., Carpenter, A. C., Nagel, S. R., Eggert, J. H., Bradley, D. K., Mackinnon, A. J., & Ping, Y. (2024). Developing time-resolved x-ray diffraction diagnostics at the National Ignition Facility (invited). Review of Scientific Instruments, 95(9). https://doi.org/10.1063/5.0219574
Vancouver
Palmer NE, Benedetti LR, Vennari CE, Nyholm PR, Petre RB, Bhandarkar N, et al. Developing time-resolved x-ray diffraction diagnostics at the National Ignition Facility (invited). Review of Scientific Instruments. 2024;95(9). doi:10.1063/5.0219574
BibTeX
@article{Palmer2024,
author = {Palmer, N E and Benedetti, L R and Vennari, C E and Nyholm, P R and Petre, R B and Bhandarkar, N and Carpenter, A C and Nagel, S R and Eggert, J H and Bradley, D K and Mackinnon, A J and Ping, Y},
title = {Developing time-resolved x-ray diffraction diagnostics at the National Ignition Facility (invited)},
journal = {Review of Scientific Instruments},
year = {2024},
volume = {95},
number = {9},
doi = {10.1063/5.0219574}
}
EUVIMED