A low-cost goniometer for contact angle measurements using drop image analysis: Development and validation

Roshan Chalise, Angash Niroula, Pooja Shrestha, Bimal Paudel, Deepak Subedi, Raju Khanal

AIP Advances · 01.08.2023 · 13 (8)

DOI: 10.1063/5.0164668 ↗

Kurz gefasst

Fachartikel in der Zeitschrift AIP Advances, online erschienen am 21.08.2023. Verfasst von Roshan Chalise und 5 weiteren Autor:innen. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Provenance Planning Commission, Lumbini, Nepal, University Grants Commission, Nepal, National Priority Project, TU, Nepal.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Contact angle measurements are a widely used method for evaluating surface wettability, but their accessibility is limited by the high cost of commercial goniometers. To address this issue, we developed a low-cost prototype goniometer using locally available materials and a smartphone-based image analysis system to measure contact angles. We validated our approach by comparing its results with those of a standard goniometer for both organic and inorganic materials. For organic samples, the accuracy is greater than 86%, and for inorganic samples, it is greater than 94%. The good agreement between the two methods demonstrates the feasibility of our proposed method for accurate contact angle measurements at a lower cost.

Förderung

  • Provenance Planning Commission, Lumbini, Nepal
  • University Grants Commission, Nepal
  • National Priority Project, TU, Nepal
  • Nepal Academy of Science, Nepal

Zitieren

APA

Chalise, R., Niroula, A., Shrestha, P., Paudel, B., Subedi, D., & Khanal, R. (2023). A low-cost goniometer for contact angle measurements using drop image analysis: Development and validation. AIP Advances, 13(8). https://doi.org/10.1063/5.0164668

Vancouver

Chalise R, Niroula A, Shrestha P, Paudel B, Subedi D, Khanal R. A low-cost goniometer for contact angle measurements using drop image analysis: Development and validation. AIP Advances. 2023;13(8). doi:10.1063/5.0164668

BibTeX
@article{Chalise2023,
  author = {Chalise, Roshan and Niroula, Angash and Shrestha, Pooja and Paudel, Bimal and Subedi, Deepak and Khanal, Raju},
  title = {A low-cost goniometer for contact angle measurements using drop image analysis: Development and validation},
  journal = {AIP Advances},
  year = {2023},
  volume = {13},
  number = {8},
  doi = {10.1063/5.0164668}
}
Kostenlos starten →