Acoustic emission detection of micro-cracks under high pressure and high temperature in a deformation large-volume apparatus at the endstation P61B, PETRA III
Review of Scientific Instruments · 01.02.2023 · 94 (2)
DOI: 10.1063/5.0107630 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 03.02.2023. Verfasst von Shuailing Ma und 2 weiteren Autor:innen. Beteiligte Einrichtung: Deutsches Elektronen-Synchrotron DESY, Notkestraße 85, 22607 Hamburg, Germany. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bundesministerium für Bildung und Forschung.
Abstract (Originalsprache)
We successfully developed an in situ acoustic emission (AE) detection setup that allows recording of AE waveforms (triggered and streaming) and simultaneous x-ray diffraction and imaging on samples deformed at high pressure and high temperature (HPHT) conditions in the Aster-15 Large Volume Press at the synchrotron beamline station P61B. This high pressure AE detection system is a powerful tool to investigate AE phenomena from the HPHT chamber. Six commercial acoustic sensors, protected by a tungsten carbide support ring on each anvil of the same material, have excellent survivability throughout each successive experiment. By pulsing each sensor in succession, the average wave velocity through the anvils and cell assembly can be determined at any press load. The distance between the sensors is obtained by x-ray radiography and by logging the positions of each hydraulic ram. This provides a basis for accurately locating AE events in the sample. The feasibility of this AE detection setup was confirmed by compression and deformation test runs using several different self-designed AE sources in specialized assemblies. The present setup proves to be extremely efficient and accurate in measuring brittle processes in samples under HPHT. It is now available for applications for beam time and experiments without x rays at P61B. Combined with synchrotron x rays, in situ pressure, temperature, strain rate and stress, and phase changes can be monitored while recording AE activity. We provide a powerful tool to investigate the origin of earthquakes, for example, causing AE emissions due to brittle dehydration reactions or phase transformations in the Earth.
Autor:innen und Institutionen
- Deutsches Elektronen-Synchrotron DESY, Notkestraße 85, 22607 Hamburg, Germany.
- Laboratoire de Géologie, CNRS - École Normale Supérieure, PSL University, 24 Rue Lhomond, 75005 Paris, France.
ORCID-Profile: Shuailing Ma, Julien Gasc, Robert Farla
Förderung
- Bundesministerium für Bildung und Forschung
Zitieren
APA
Ma, S., Gasc, J., & Farla, R. (2023). Acoustic emission detection of micro-cracks under high pressure and high temperature in a deformation large-volume apparatus at the endstation P61B, PETRA III. Review of Scientific Instruments, 94(2). https://doi.org/10.1063/5.0107630
Vancouver
Ma S, Gasc J, Farla R. Acoustic emission detection of micro-cracks under high pressure and high temperature in a deformation large-volume apparatus at the endstation P61B, PETRA III. Review of Scientific Instruments. 2023;94(2). doi:10.1063/5.0107630
BibTeX
@article{Ma2023,
author = {Ma, Shuailing and Gasc, Julien and Farla, Robert},
title = {Acoustic emission detection of micro-cracks under high pressure and high temperature in a deformation large-volume apparatus at the endstation P61B, PETRA III},
journal = {Review of Scientific Instruments},
year = {2023},
volume = {94},
number = {2},
doi = {10.1063/5.0107630}
}
EUVIMED