Retrospective Benchmarking of CAD EYE and ColoMAIA-2 Using Delay-Based Sensitivity and False-Alarm Metrics

J Kral, K Knotkova, M Hradis, P Musil, P Svoboda, J Svet

Endoscopy · 01.06.2026 · 58 (S 03 S 03) · S375

DOI: 10.1055/s-0046-1821294 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.06.2026. Verfasst von J Kral und 5 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

To establish objective metrics for retrospective CADe evaluation and use them to directly compare Fujifilm CAD EYE with Maia Labs ColoMAIA-2 on identical colonoscopy recordings.

Methods

This retrospective single-center study analyzed video data from 150 colorectal cancer screening colonoscopies (1945 minutes) performed by three expert endoscopists (>10 years’ experience). A total of 189 clinically significant polyps were identified in the recordings. CAD EYE operated in real time during procedures, while ColoMAIA-2 was applied offline. Outcome measures included: 1) detection delay from first polyp appearance; 2) delay-specific sensitivities at 100–5000 ms; 3) False alarms, expressed as false detections per minute; 4) percentage of video time with false alarms. False alarms were assessed only during active mucosal inspection (excluding polyp examinations and polypectomies). Bootstrap resampling was used to compute confidence intervals and p-values.

Results

Across all short delays, ColoMAIA-2 achieved substantially higher sensitivity than CAD EYE. At 100 ms, sensitivity was 47.6% vs. 16.9%; at 200 ms, 67.2% vs. 33.9%; at 300 ms, 78.8% vs. 45.5%. At the clinically meaningful 500-ms delay, ColoMAIA-2 reached a sensitivity 86.8% (CI 81.5–91.5) compared with 62.4% (CI 55.6–69.3) for CAD EYE (p

Zitieren

APA

Kral, J., Knotkova, K., Hradis, M., Musil, P., Svoboda, P., & Svet, J. (2026). Retrospective Benchmarking of CAD EYE and ColoMAIA-2 Using Delay-Based Sensitivity and False-Alarm Metrics. Endoscopy, 58(S 03 S 03), S375. https://doi.org/10.1055/s-0046-1821294

Vancouver

Kral J, Knotkova K, Hradis M, Musil P, Svoboda P, Svet J. Retrospective Benchmarking of CAD EYE and ColoMAIA-2 Using Delay-Based Sensitivity and False-Alarm Metrics. Endoscopy. 2026;58(S 03 S 03):S375. doi:10.1055/s-0046-1821294

BibTeX
@article{Kral2026,
  author = {Kral, J and Knotkova, K and Hradis, M and Musil, P and Svoboda, P and Svet, J},
  title = {Retrospective Benchmarking of CAD EYE and ColoMAIA-2 Using Delay-Based Sensitivity and False-Alarm Metrics},
  journal = {Endoscopy},
  year = {2026},
  volume = {58},
  number = {S 03 S 03},
  pages = {S375},
  doi = {10.1055/s-0046-1821294}
}
Kostenlos starten →